4
Potential Device Interference.
Bluetooth and Wireless LAN devices operate within the same
radio frequency range and may interfere with one another.
If you are using the AortaScan AMI 9700 Bluetooth link and wireless LAN devices
simultaneously, you may experience less‑than‑optimal network performance or even lose
your network connection. If this happens, you may need to move the AortaScan AMI 9700
and ScanPoint
®
host computer to an area away from the 2.4 GHz wireless LAN devices (40
meters/44 yards, or more).
CAUTION
Medical electrical equipment requires special precautions regarding electromagnetic
compatibility (EMC) and must be installed and operated according to the instructions in this
manual. For more information, see the
section on page 72.
To maintain electromagnetic interference (EMI) within certified limits, the AortaScan AMI 9700
system must be used with the cables, components, and accessories specified or supplied
by Verathon
®
. For additional information, see the
System Components & Accessories
and
sections. The use of accessories and/or cables other than those
specified or supplied may result in increased emissions and/or decreased immunity of the
system.
The AortaScan AMI 9700 system should not be used adjacent to or stacked with other
equipment. If adjacent or stacked use is necessary, the system should be observed to verify
normal operation in the configuration in which it will be used.
This device can radiate radio frequency energy and is very unlikely to cause harmful interference
with other devices in the vicinity. There is no guarantee that interference will not occur in a
particular installation. Evidence of interference may include degradation of performance in
this device or other devices when operated simultaneously. If this occurs, try to correct the
interference by using the following measures:
• Turn devices on and off in the vicinity to determine the source of interference
• Reorient or relocate this device or other devices
• Increase the separation between devices
• Connect the device to an outlet on a circuit different than the other device(s)
• Eliminate or reduce EMI with technical solutions (such as shielding)
• Purchase medical devices that comply with IEC 60601‑1‑2 EMC Standards
Be aware that portable and mobile radio frequency communications equipment (cellular phones,
etc.) may affect medical electrical equipment; take appropriate precautions during operation.
CAUTION