68
PRODUCT SPECIFICATIONS
COMPONENT SPECIFICATIONS
CONSOLE & PROBE SPECIFICATIONS
Table 20. General Specifications
ITEM
SPECIFICATION
Input
Lithium‑ion battery.
Output
No load to full load at rated voltage. Refer to unit label.
Insulation
The power supply is Class I with basic insulation to each terminal.
Transient overvoltage
Category II
Weight
5.2 lb (2.36 kg) (with battery)
Display
13.36 cm W x 10.13 cm H (5.26 in W x 3.99 in H) (640 x 480 pixels, 120 dpi)
Integrated printer
Thermal printer
Table 21. Ultrasound Acoustic Output Parameters
Values in this table are the maximum readings obtained from three test results.
ACOUSTIC OUTPUT
MI
I
SPTA.3
(mW/cm
2
)
I
SPPA.3
(W/cm
2
)
Global Maximum
Value
0.519*
0.632
9.35
Associated
Acoustic
Parameter
p
r.3
(MPa)
0.684
W
o
(mW)
1.55
1.44
f
c
(MHz)
1.74
1.74, 2.63
†
1.74
z
sp
(cm)
1.90
1.90
Beam
dimensions
x
‑6
(cm)
0.321
y
‑6
(cm)
0.334
PD
(µsec)
2.93
2.93
PRF
(Hz)
400
400
EDS
Az. (cm)
7.40, 7.38
†
Ele. (cm)
7.40, 7.38
†
TIS/TIB/TIC range
0.0‑1.0*
*
Both MI and TI values are below 1.0.
†
Each scan point along the scan line consists of two transmit pulses. The first pulse is 1 cycle at 2.95 MHz and the second pulse is 5
cycles at 1.74 MHz. Data for each pulse is provided and separated by a comma.
Содержание AORTASCAN AMI 9700
Страница 1: ...AORTASCAN AMI 9700 Operations Maintenance Manual...
Страница 2: ...0900 4411 04 60...
Страница 8: ......
Страница 87: ......
Страница 88: ......