ELGA
MEDICA 7/15 US
Operator Manual
MEDICA 7/15 US Version 2
– 10/15
Page 25
7.2 Replacing LC141 Ion-exchange Cartridge Pack
The Ion-exchange cartridge pack should be replaced in the following
circumstances:
The water purity alarm monitor indicates that water purity has
fallen below limits and the pack requires changing.
If the system is being re-commissioned or sanitized after an
extended period in which it was not used.
When indicated by the change reminder.
Immediately before and after use of the emergency by-pass
loop.
Step 1 - Switch Unit Off
1. SWITCH the
MEDICA
off at the power switch at
the top left hand side of the unit.
2. RELIEVE any residual pressure from the system,
by waiting several minutes before proceeding.
Then slowly opening the bleed nipple on the
0.2µm capsule filter until water ceases, use an
absorbent cloth to soak up the water and re-
tighten the bleed nipple.
WARNING!
ENSURE THE UNIT IS ISOLATED BEFORE
REMOVING THE ION-EXCHANGE
CARTRIDGE.
Step 2 - Remove Ion-exchange Cartridge
1. OPEN the front door.
2. PUSH on cartridge top cap.
3. LIFT up cartridge.
4. SLIDE out cartridge.
5. DISCARD used Ion-exchange cartridge.
Note:
The consumable is non-hazardous.
Dispose of as ordinary waste, observing
all local and national regulations.
Step 3 - Replace Ion-exchange Cartridge
1. REMOVE a new cartridge pack from its
packaging.
2. REMOVE the sealing plugs from the inlet and
outlet ports.
3.
WET ‘O’ rings and SLIDE new cartridge into
position.
4. POSITION cartridge onto spigots at rear, PUSH
into unit.
5. ENSURE guide has dropped down past retainer.
6. CLOSE front door.
7. RESET Ion-exchange pack replacement timer.
(See Section 5.4 - Setting up Display and
Replacement Timer)
.
8. Follow steps 14 - 23 of
Section 5.6
Initial Start up.
Ion-exchange
cartridge
Removal and Fitting of Ion-exchange
Cartridge
Removing
Replacing
Ion-exchange Cartridge
Bleed
nipple
0.2µm
capsule filter
0.2µm Capsule filter & Bleed Nipple