Operator Manual
MEDICA 7/15 US Version 2
– 10/15
Page 13
Drain
Outlet to
reservoir
Feed to
Analyzer
Delivery
pump feed
Delivery Pump and Drain Connections
5.3 Connecting up the
MEDICA
Once the
MEDICA
unit has been positioned either on a wall or on a
bench, it should be connected as follows:
Mains water inlet tube
Drain
Feed to analyzer
Outlet to reservoir
Delivery Pump Feed
Step 1 - Fitting Tubes
1. PUSH in collet on connector.
2. PULL out transit plug.
3. CUT a clean square end on an 8mm (5/16") O/D
semi rigid tube.
4. PUSH tube into connector.
CAUTION!
Do not restrict drain line.
CAUTION!
If the water supply is at a pressure greater
than 6 bar (90 psi) fit a pressure regulator
(LA512).
Connector
Collet
Blanking
Transit Plug
Pull Plug
Depress
Collet
Clean Square
Cut End
Tube 8mm OD
Simply push in tube to attach
Tube in secured position
Fitting Tubes
Feed Water Inlet Connections
Emergency
by-pass
connection
Blanking
plug