![Veolia ELGA MEDICA 15 BP-US Скачать руководство пользователя страница 22](http://html1.mh-extra.com/html/veolia/elga-medica-15-bp-us/elga-medica-15-bp-us_operators-manual_1003042022.webp)
MEDICA 7/15
Operator Manual
ELGA
Page 16
MEDICA 7/15 US Version 2
– 10/15
Step 4 - Water Purity Unit Setting
This screen allows preferred water purity unit of measure to be
set, to either,
M
.cm
or
µS/cm
. Once selected, all future water
purity measurements will be displayed with your unit of choice.
Select the option required using the
button and accept with
the
button.
Step 5 - Analyzer Delivery Flow Adjustment
The
MEDICA 7/15
includes the ability to adjust the delivery flow
rate to match the needs of specific analyzers.
Select the flow-curve from Graph A, Delivery Flows. Please
contact your analyzer supplier if you are unsure of your analyzer
requirements or alternatively contact ELGA Technical Support.
Select the flow required along the ‘Y’ axis, Flow (ml/min) and
draw a line parallel
with the ‘X’ axis. The line will converge with
a flow curve. Draw a line parallel with the ‘Y’ axis. Confirm that
the pressure indicated is suitable for the analyzer.
The flow curve is identified as a percentage i.e. 80%.
Select the percentage (flow curve) using the
button and
accept with the
button.
MEDICA
Flow (ml/min)
Pressure (psi)
Graph A - Delivery Flows
Megaohm
Microsiemen
Water Purity Settings
100%
Analyzer Delivery Flow
Adjustment
Содержание ELGA MEDICA 15 BP-US
Страница 1: ...MEDICA 7 15 US Operator Manual...
Страница 2: ......
Страница 4: ...MEDICA 7 15 Operator Manual ELGA Page ii MEDICA 7 15 US Version 2 10 15...