ELGA
MEDICA 7/15 US
Operator Manual
MEDICA 7/15 US Version 2
– 10/15
Page 23
7. MAINTENANCE
Any maintenance work not detailed in this handbook should be carried
out by an approved supplier or distributor. If further information is
required on any aspect of maintenance please contact Customer
Service.
Identification of Consumables
There are four types of unique replacement consumables designed for
use in the
MEDICA
units and these are illustrated with the following
part numbers:
LC140 Pre-treatment cartridge
LC141 Ion-exchange cartridge
LC105 UV lamp
LC125 0.2µm capsule filter
All consumables are accessible after opening the front door cover.
To protect the inlet solenoid valve, RO boost pump (when fitted) and
Delivery pump from possible debris in the water, the unit incorporates
two strainers.
WARNING!
ALWAYS
CHECK
THAT
THE
MAINS
ELECTRICAL POWER AND FEED WATER
SUPPLIES ARE SWITCHED OFF BEFORE
ATTEMPTING TO CHANGE THE
MEDICA
CONSUMABLES.
Frequency of Consumable
Replacement
The following frequency of consumable replacement is recommended
as a guide assuming typical usage.*
Pre-treatment
- LC140
maximum 6 months
Ion exchange
- LC141
maximum 6 months**
UV lamp
- LC105
maximum 12 months
0.2µm capsule filter
- LC125
maximum 6 months
Reverse Osmosis
- LC143
every 2 - 3 years
(not an operator
replacement item)
*
These frequencies are only estimates and replacement will
depend on the application and feed water quality.
**
Standard conditions are 4 bar (60 psi) inlet pressure at
15°C, potable water with clean pre-filter.
CAUTION!
Ensure that the display and replacement
timer settings are reset after replacing
consumables. (Refer to Section 5.4).
LC140
Pre-treatment
cartridge
LC141
Ion-exchange
cartridge
LC105
UV lamp
0.2µm capsule filter
MEDICA
Replacement Consumables