MEDICA 7/15
Operator Manual
ELGA
Page 20
MEDICA 7/15 US Version 2
– 10/15
17. SLOWLY OPEN the upper bleed nipple on the
0.2µm capsule filter until water, free of air
bubbles, flows out - use an absorbent cloth to
prevent splashing of internal components.
18. Allow the system to operate until the water purity
indicator displays a value within acceptable
operating values.
19. PRESS the process button to stop the unit.
20. RE-CONNECT the feed to the analyzer.
21. PRESS the process button. The unit will now run
automatically and fill the treated water reservoir.
22. When the analyzer calls for water (i.e. its inlet
valve is open) the delivery pump and UV will
operate.
23. CHECK the unit for leaks.
24. The unit is now ready for operation.
Bleed
nipple
0.2µm
capsule filter
Bleed Nipple