ELGA
CENTRA – MDS/LDS/RDS US Operator Manual
CENTRA – MDS/LDS/RDS US Version 01 – 08/14
Page 27
Step 19 - Data transmit
The frequency of the data transmittal to the printer or PC is
selectable.
1.
PRESS SCROLL
to change transmit intervals
(1,5,15,30min / 1,6 hour)
OR
PRESS TICK
to proceed to step 20.
2.
PRESS TICK
.
Note: PRESS PRINT during normal operation and data
will be transmitted to the recording device.
Step 20 - Remote control station selection (node
addressing)
Each remote control station connected to the
CENTRA
must be registered for its commands to be recognized.
Once registered the remote control station can be used
to adjust functionality and monitor alarms locally.
The viewing angle of each remote control station is
specific to its mounting position and should be adjusted
as described in Section 5.2, Step 16.
1.
PRESS RESET
OR
PRESS TICK
to complete settings (User
PASSkey)
OR
PRESS TICK
to proceed to step 21 (Master
PASSkey only).
Note: At this stage the CENTRA screen may timeout
before each remote display are registered. This
does not affect registration of the remote display.
2.
ENSURE the RS485 network is correctly installed
and terminated as described in ‘ELGA LabWater
Installation Guide - RS485 Local area Network
design considerations and installation guide’
(Available from ELGA LabWater Technical Support).
3.
POWER ON remote control stations.
4.
PRESENT user PASSkey (any colour) at each
remote display in the order that you wish them to be
registered.
Note: Re-enter the user menu and return to step 20 to
check the registration of the remote displays.
Data transmit screen
Remote control station
selection screen