3
5. Maintenance
No routine or extraordinary maintenance is required except periodic cleaning.
Cleaning
Disconnect the unit from the power supply and use a cloth dampened with a non-flammable non-aggressive detergent.
Repair
Repairs must be carried out by authorized Velp personnel only.
The transport of the instrument by freight forwarders, couriers or others must be carried out using the original shockproof
packaging. Follow any instructions on the original packaging (e.g. palletizing).
It is the responsibility of the user to properly decontaminate the unit in case of hazardous substances remaining on the
surface or interior of the device. If in doubt about the compatibility of a cleaning or decontamination product, contact the
manufacturer or distributor.
6. Technical data
AREX 5
ARE 5
RC 5
General
features
Model
F20500570
F20510570
F20500560
F20510560
F20700610
F20710610
Voltage
230 V
50/60 Hz
115 V
60 Hz
230 V
50/60 Hz
115 V
60 Hz
230 V
50/60 Hz
115 V
60 Hz
Power input
630 W
630 W
600 W
Current consumption
2,8 A
5,5 A
2,8 A
5,5 A
2,6 A
5,2 A
Dimensions (WxHxD)
160x270x85 mm (6,3x10,6x3,3 in)
Weight
1,7 kg (3,7 lb)
1,4kg (3,1 lb)
Construction material
Aluminum body
–Technopolymer enclosure
Working in continuous
Admitted
Maximum load on the
plate
25 kg
Noisiness
<< 80 dBa
Environmental
temperature admitted
+
5…+40 °C
Storage temperature
admitted
-10
…+60 °C
Level of electrical
protection CEI
EN60529
IP 42
Max humidity
80%
Overvoltage category
II
Pollution degree CEI
EN61010-1
2
Max altitude
4000 m
Heating
Heating plate power
output
600 W
Heating plate
dimensions
Ø 135 mm
Temperature range
50 ÷ 310 °C
Heat control type
Analogic
Heating plate
construction material
CerAlTop
™
Aluminum
Safety circuit
Separated with dedicated probe
Hot plate alarm
Over 50°C
Overtemperature
alarm
Over 330°C