T
he following table shows the codes and text messages of the status
message
"
F
unction check
"
and provides information on causes as well
as corrective measures
.
C
ode
T
ext mes
-
sage
C
ause
R
emoval
C
700
S
imulation
l
S
imulation active
l
F
inish simulation
l
W
ait for the automatic end
after
60
mins
.
T
he following table shows the codes and text messages of the status
message
"
O
ut of speci
fi
cation
"
and gives instructions for the reason
and removal
.
C
ode
T
ext mes
-
sage
C
ause
R
emoval
S
600
U
npermissible
electronics
temperature
l
T
emperature of the elect
-
ronics in the not speci
fi
ed
section
l
C
heck ambient temperatu
-
re
l
I
solate electronics
l
U
se instrument with higher
temperature range
T
he following table shows the codes and text messages of the status
message
"
M
aintenance
"
and provides information on causes as well
as corrective measures
.
C
ode
T
ext mes
-
sage
C
ause
R
emoval
M
500
E
rror with the
reset delivery
status
l
W
ith the reset to delivery
status
,
the data could not
be restored
l
R
epeat reset
l
L
oad
XML
fi
le with sensor
data into the sensor
M
501
E
rror in the
non
-
active li
-
nearization
table
l
H
ardware error
EEPROM
l
E
xchange the electronics
l
S
end instrument for repair
M
502
E
rror in the
diagnosis me
-
mory
l
H
ardware error
EEPROM
l
E
xchange the electronics
l
S
end instrument for repair
F
unction check
O
ut of speci
fi
cation
M
aintenance
(
mainte
-
nance requirement
)
VEGAPULS
62
•
4
…
20
m
A
/
HART
two
-
wire
49
9
D
iagnosis and service
36503
-
EN
-
100607