55
11 Supplement
VEGABAR 81 • 4
… 20 mA
45025-EN-140228
Dead time
≤ 45 ms
Rise time
≤ 35 ms (10 … 90 %)
Step response time
Ʋ
VEGABAR 81
≤ 80 ms (ti: 0 s, 10 … 90 %)
Ʋ
VEGABAR 81 - IP
68
(25
bar)
≤ 200 ms (ti: 0 s, 10 … 90 %)
To this amounts the reaction time of the isolatng system. This time varies from values < 1 s with
compact chemical seals to several seconds with capillary systems.
Example: Flange-type chemical seal DN 80, filling silicone oil KN 2.2, capillary length 10 m, meas
-
uring range 1 bar
Process temperature
Reaction time approx.
+40
°C (+104
°F)
+20
°C (+58
°F)
ca. 3 s
-20
°C (-4
°F)
ca. 11 s
Damping (63 % of the input variable)
0 … 999 s, adjustable
Reference conditions and actuating variables (according to DIN EN 60770-1)
Reference conditions according to DIN EN
61298-1
Ʋ
Temperature
+18 … +30
°C (+64
… +86 °F)
Ʋ
Relative humidity
45
…
75
%
Ʋ
Air pressure
860 … 1060 mbar/86 … 106
kPa (12.5
…
15.4
psi)
Determination of characteristics
Limit point adjustment according to IEC
61298-2
Characterstic curve
Linear
Reference installation position
upright, diaphragm points downward
Influence of the installation position
depending on the chemical seal version
Deviation on the current output due to
strong, high frequency electromagnetic
fields acc. to EN 61326
<
±150
µA
Deviation (according to IEC 60770)
Applies to the
digital
signal output (HART, Profibus PA, Foundation Fieldbus) as well as to the
analogue
current output 4
… 20 mA and refers to the set span. Turn down (TD) is the ratio nominal
measuring range/set span.
Accuracy class
Non-linearity, hysteresis and re-
peatability with TD 1 : 1 up to 5 : 1
Non-linearity, hysteresis and re-
peatability with 5 : 1
0.2 %
< 0.2 %
<
0.04
% x TD
Influence of the product or ambient temperature
Thermal change zero signal and output span
Applies to the
digital
signal output (HART, Profibus PA, Foundation Fieldbus) as well as to the
analogue
current output 4
… 20 mA and refers to the set span. Turn down (TD) is the ratio nominal
measuring range/set span.