54
11 Supplement
VEGABAR 81 • 4
… 20 mA
45025-EN-140228
Output variable
Output signal
4
… 20 mA
Range of the output signal
3.55
… 22.0 mA (default setting)
Signal resolution
0.3
µA
Failure signal current output (adjustable)
Last valid measured value, ≥ 21 mA, ≤ 3.6 mA
Max. output current
21.5
mA
Load
see load under Power supply
Damping (63 % of the input variable),
adjustable
0 … 999 s
Indication value - DIsplay and adjustment module
1)
Ʋ
Displayed value 1
Pressure in bar/mbar
Ʋ
Displayed value 2
Pressure in bar/mbar
Output variable - Second current output
For details on the operating voltage see chapter "Voltage supply"
Output signal
4
… 20 mA (passive)
Range of the output signal
3.8 …
20.5
mA (default setting)
Signal resolution
0.3
µA
Failure signal current output (adjustable)
Last valid measured value, ≥ 21 mA, ≤ 3.6 mA
Max. output current
21.5
mA
Starting current
≤ 10 mA for 5 ms after switching on, ≤ 3.6 mA
Load
Load resistor, see chapter "Voltage supply"
Damping (63 % of the input variable),
adjustable
0 … 999 s
Dynamic behaviour output
Dynamic characteristics depending on medium and temperature
90 %
100 %
10 %
t
t
T
A
t
t
S
2
1
Fig. 28: Sudden change of the process variable. t
T
: dead time; t
A
: rise time; t
S
: jump response time
1 Process variable
2 Output signal
1)
The indication values can be assigned individually