42
6 Set up with the display and adjustment module
VEGABAR 81 • 4
… 20 mA
45025-EN-140228
If the instrument is equipped with a display and adjustment module,
the data in the sensor can be saved in the display and adjustment
module. The procedure is described in the operating instructions
manual "
Display and adjustment module
" in the menu item "
Copy
sensor data
". The data remain there permanently even if the sensor
power supply fails.
The following data or settings for adjustment of the display and ad
-
justment module are saved:
•
All data of the menu "
Setup
" and "
Display
"
•
In the menu "
Additional adjustments
" the items "
Sensor-specific
units, temperature unit and linearization
"
•
The values of the user programmable linearization curve
The function can also be used to transfer settings from one instru
-
ment to another instrument of the same type. If it is necessary to
exchange a sensor, the display and adjustment module is inserted
into the replacement instrument and the data are likewise written into
the sensor via the menu item "
Copy sensor data
".