DSGH Installation and Operation Guide
67
Advanced functions
Temperature Test
This mode manually forces the sensor’s temperature probe output to a
specified value. This is useful for verifying the scintillator sensor
temperature compensation.
Procedure 4.9: To perform a temperature test
Selecting the transmitter’s type and location
Gauge Setup tab
1.
Select
Diagnostics | Test | Temperature Test
.
2.
Click
Enter
.
3.
Remove the gauge from control. Enter the value of the new
temperature to force.
4.
Click
OK
.
5.
The transmitter functions in this mode until it times out (1
hour), or you click
Exit
and
OK
.
Содержание DSGH
Страница 2: ......
Страница 4: ...2 DSGH Installation and Operation Guide NOTES ...
Страница 8: ...6 DSGH Installation and Operation Guide NOTES ...
Страница 12: ...10 DSGH Installation and Operation Guide Preface NOTES ...
Страница 32: ...30 DSGH Installation and Operation Guide Installation Mounting the DSGH gauge ...
Страница 56: ...54 DSGH Installation and Operation Guide Calibration NOTES ...
Страница 94: ...92 DSGH Installation and Operation Guide Diagnostics and repair NOTES ...
Страница 98: ...96 DSGH Installation and Operation Guide NOTES ...
Страница 99: ......