DSGH Installation and Operation Guide
65
Advanced functions
Auxiliary Input Test
This mode simulates the auxiliary input frequency at a user-defined number
of counts. The effect of auxiliary input counts depends on the auxiliary input
mode.
Examples:
• Temperature probe
• Flow meter
• Second transmitter
While in this mode, after entering a number of counts, it may be useful to
look at the
Process Chain
tab to view the variables affected by the auxiliary
input counts value.
Procedure 4.7: To perform an auxiliary input test
3.
Remove the gauge from control. Enter the value of the new
counts to force.
4.
Click
OK
.
The transmitter functions in this mode until it times out (1
hour), or you click
Exit
and
OK
.
1.
Select
Diagnostics | Test | Auxiliary Input Test
.
2.
Click
Enter
.
3.
Remove the gauge from control. Enter the auxiliary counts.
4.
Click
OK
.
The transmitter functions in this mode until it times out (1
hour), or you click
Exit
and
OK
.