Advanced functions
New Hardware tab
Responding to the New hardware found message
When new hardware is installed
When you install a new CPU board or sensor assembly, you must verify
installation in Ohmview 2000 to enable new backups of the EEPROMs.
Procedure 4.3: To verify the “New Hardware Found” message
When new hardware is not installed
If the error message New hardware found appears, an EEPROM is probably
corrupt.
The messages “CPU EEPROM Corrupt” or “Sensor EEPROM Corrupt” may
also appear in the history.
1.
Select
Diagnostics | New hardware | New CPU
or
New Sensor
.
2.
Click
OK
.