DSGH Installation and Operation Guide
Calibration
Procedure 3.2: To choose an equation linearizer type
Checking the gauge repeatability
Check the gauge’s measurement repeatability before performing the
calibration.
To check the repeatability of the sensor, perform a data collection 3 – 4
times on the same sample. If the sensor counts vary widely, you should
increase the Data collection interval parameter.
Perform a data collection to enable simple measurement of the process,
without altering the calibration or standardization values. It lets the system
measure the process and report the number of sensor counts.
1.
Select
Setup | Gauge Setup | Linearizer Type
.
2.
Click
Equation
.