FX8x Series User Manual
Page 17 of 50
Selecting a wavelength for PM measurement
Select a wavelength by pressing the
|Up|
and
|Down|
buttons, then press the
|Shift/Enter|
button to confirm the choice. The new wavelength then appears in the top right corner of the
display.
For the previous wavelength to remain effective, select a wavelength and then press
the
|F1| (EXIT)
button
.
If the fiber under test carries a signal with WaveID details, select
[ AUTO]
. Then, the device
(FX82, FX84, or FX85) determines the wavelength automatically.
If the [ AUTO] option for a signal without WaveID or modulated is selected, the λ
value goes blank, and the measurement results are NOT valid.
6.2.1 Setting Power Reference Level
To measure the fiber under test span loss, first set a reference level for each test wavelength.
In the
PM1/PM2
mode, press the
|F3|
button
(MORE)
and then press
|F2| (REF)
.