FX8x Series User Manual
Page 32 of 50
ORL calibration in progress
4. Connect the reference cable to the ORL port. The reference cable is provided with the
device.
5. Press
|F3| (REF_C)
, then set the -14.00 value by pressing
|Up|
and
|Down|
.
Setting ORL reference value
6. After setting the
-14.00
value press
|Shift/Enter|
. The reference value is then saved to
the device memory.
7. Disconnect the reference cable and press
|F1| (CALIBR)
again. The
CALIBRATION
notification is shown onscreen for several seconds again.
Now the device is ready for ORL measurement.