Note: A 25 W D-sub to 4 W BNC adaptor block is available through your VBOX distributor, part number
RLVBACS054
Using VBOX Setup software, logging of the analogue channel data can be switched on or off. It is also possible to
change the name of each input channel and configure scale and offset values for calibration of sensors. A scale value of
1 and offset of 0 correspond to a channel reading in V DC. This means that the value stored on the compact flash card
for the channel will also be in volts. When using a sensor such as a load cell, it may be desirable to store a reading in
kg. In this case, changing the scale and offset to suit the sensor data sheet allows the data stored onto the compact
flash to be in kg. When changing settings for an analogue channel using VBOX Setup software, a live data view of the
current channel is shown. The value shown is the value after scale and offset is applied and can therefore be used to aid
sensor calibration.
Note: The 5 V regulated output on pin16 is only good for VBOX power supply voltages >8.5 V
https://en.racelogic.support//Product_Info/VBOX_Data_Loggers/VBOX_3i_Range/VBOX_3i_User_Guide_(All_Variants)/09_-
_VB3i_Analogue_and_Digital_I%2F%2FO