21
on operating the blood
pressure monitor and repeat
measurement accordingly.
Err 6
N/A
SmartCheck™
mode error
Multiple errors detected
during measurement in
SmartCheck™ mode,
preventing a reliable final
result. Refer to chapter 5 on
operating the blood pressure
monitor and repeat
measurement accordingly.
HI
N/A
Pulse or cuff
pressure too
high
Cuff pressure is too high (over
300 mmHg) or pulse detected
is too high (over 200 beats per
minute). Pause and relax for at
least 5 minutes and repeat
measurement.
LO
N/A
Pulse too low
Pulse detected is too low (less
than 40 beats per minute).
Pause and relax for at least 5
minutes and repeat
measurement
Troubleshooting
Problems
Solutions
The display remains blank when the
device is switched on.
1. Make sure that the batteries have
been installed correctly according to
the indicated polarity.
2. If batteries are drained, replace
with new batteries.
The device fails to measure blood
pressure values, or the measured
values are too low or too high.
1. Make sure that the cuff has been
applied and positioned properly.
2. Pause for at least 5 minutes.
Repeat measurement and remain
relaxed, still and do not talk during