Chapter 6 Error messages and troubleshooting
If an error occurs during a measurement, the
measurement is discontinued and a corresponding error
code is displayed (Example: Error No. 1).
Error
Icon shown
on display
Description
Possible cause and solution
Err
1
N/A
Signal too
weak
The pulse signals detected on
the cuff are too weak.
Reposition/readjust the cuff
and repeat the measurement.
Err
2
Movement
detected
During measurement,
movement or muscle tension
is detected by the cuff. Repeat
measurement and remain
relaxed, still and do not talk
during measurement.
Err 3
Cuff pressure
too low
An adequate pressure cannot
be generated in the cuff. It can
be due to a leak, loose cuff or
cuff tubing incorrectly
connected. Check cuff tubing
connection and
reposition/readjust the cuff.
Repeat measurement.
Err 5
N/A
Abnormal
result
The measuring signals are
inaccurate or unstable,
therefore no result can be
displayed. Refer to chapter 5