3. Warnings and Precautions
PCT-90LH User Manual
13
EN
G
LISH
3.2
Electricity-related Safety Precautions
To avoid risk of electric shock, this equipment must only be connected
to supply mains with protective earth.
Check the status of the power source, PC and cables prior to operating the
equipment.
Ensure that
Main Power Switch
is set to off when the equipment is not in use.
Always disconnect the power supply before cleaning the equipment.
Always keep electrical cords away from hot appliances or radiators.
DO NOT place the PC or peripheral equipment connected to the PC in the
immediate vicinity of the patient.
The equipment and PC should be connected to a common protective earth.
Never overload the equipment’s circuit by sharing it with too many appliances.
Use the same power circuit for the PC and the equipment.
Combining this equipment with other devices
Do not connect this equipment to devices which are not designated as a part of
the system.
Do not connect this equipment to a Multiple Portable Socket-Outlet (MPSO) or
extension cord which is not provided with the equipment.
Electromagnetic compatibility
This equipment complies with IEC standard 60601-1-2.
Medical electrical equipment is subject to special Electromagnetic Compatibility
(EMC) preventive measures. It must be installed and operated as specified in
EMC information.
If high-voltage systems, radio link systems or MRI systems are located within 5 m
of the unit, please observe the specifications stated in the installation
requirements.
Portable Radio Frequency (RF) communications equipment may interfere with
medical electrical equipment. Therefore, the use of mobile wireless phones in
medical offices or hospital environments must be prohibited.
For more details, refer to
15.3 Electromagnetic Compatibility (EMC)
Information
.
Please also observe the Electro-Static Discharge (ESD) protective measures
described.
Содержание Premium PAX-i3D
Страница 1: ......
Страница 2: ...PCT 90LH User Manual 3...
Страница 27: ...4 Imaging System Overview PCT 90LH User Manual 21 ENGLISH 4 4 Imaging System Configuration...
Страница 29: ...4 Imaging System Overview PCT 90LH User Manual 23 ENGLISH 4 5 Equipment Overview...
Страница 44: ...4 Imaging System Overview 38 PaX i3D Green Premium User Manual Left blank intentionally...
Страница 52: ...5 Imaging Software Overview 46 PaX i3D Green Premium User Manual Left blank intentionally...
Страница 58: ...6 Getting Started 52 PaX i3D Green Premium User Manual Left blank intentionally...
Страница 102: ...8 Acquiring i CEPH Images Optional 96 PaX i3D Green Premium User Manual Left blank intentionally...
Страница 122: ...9 Acquiring Dental CT Images 116 PaX i3D Green Premium User Manual Left blank intentionally...
Страница 146: ...11 Acquiring 3D PHOTOs Optional 140 PaX i3D Green Premium User Manual Left blank intentionally...
Страница 148: ...12 Troubleshooting 142 PaX i3D Green Premium User Manual Left blank intentionally...
Страница 152: ...13 Cleaning and Maintenance 146 PaX i3D Green Premium User Manual Left blank intentionally...
Страница 154: ...14 Disposing of the Equipment 148 PaX i3D Green Premium User Manual Left blank intentionally...
Страница 161: ...15 Technical Specifications PCT 90LH User Manual 155 ENGLISH Maximum Rating Charts Emission Filament Characteristics...
Страница 166: ...15 Technical Specifications 160 PaX i3D Green Premium User Manual Left blank intentionally...
Страница 189: ...16 Appendices PCT 90LH User Manual 183 ENGLISH...
Страница 204: ......