NOTE 1) At 80MHz and 800MHz, the separation distance for the higher frequency
range applies.
NOTE 2) These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects, and
people.
a. Field strength from fixed transmitters, such as base stations for radio
(cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM
radio broadcast, and TV broadcast cannot be predicted theoretically with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in
the location in which the
PCT-90LH
is used exceeds the applicable RF compliance
level above, the
PCT-90LH
should be observed to verify normal operation. If
abnormal performance is observed, additional measures may be necessary, such as
re-orienting or relocating the
PCT-90LH
b. Over the frequency range 150kHz to 80MHz, field strengths should be less than
[V
1
] V/m.
Immunity and Compliance Level
Immunity test
IEC 60601-1-2
Test level
Actual Immunity
Level
Compliance
Level
Conducted RF
IEC 61000-4-6
3Vrms
150kHz to 80MHz
3Vrms
3Vrms
Radiated RF
IEC 61000-4-3
3Vrms
80MHz to 2.5GHz
3V/m
3V/m
Содержание Premium PAX-i3D
Страница 1: ......
Страница 2: ...PCT 90LH User Manual 3...
Страница 27: ...4 Imaging System Overview PCT 90LH User Manual 21 ENGLISH 4 4 Imaging System Configuration...
Страница 29: ...4 Imaging System Overview PCT 90LH User Manual 23 ENGLISH 4 5 Equipment Overview...
Страница 44: ...4 Imaging System Overview 38 PaX i3D Green Premium User Manual Left blank intentionally...
Страница 52: ...5 Imaging Software Overview 46 PaX i3D Green Premium User Manual Left blank intentionally...
Страница 58: ...6 Getting Started 52 PaX i3D Green Premium User Manual Left blank intentionally...
Страница 102: ...8 Acquiring i CEPH Images Optional 96 PaX i3D Green Premium User Manual Left blank intentionally...
Страница 122: ...9 Acquiring Dental CT Images 116 PaX i3D Green Premium User Manual Left blank intentionally...
Страница 146: ...11 Acquiring 3D PHOTOs Optional 140 PaX i3D Green Premium User Manual Left blank intentionally...
Страница 148: ...12 Troubleshooting 142 PaX i3D Green Premium User Manual Left blank intentionally...
Страница 152: ...13 Cleaning and Maintenance 146 PaX i3D Green Premium User Manual Left blank intentionally...
Страница 154: ...14 Disposing of the Equipment 148 PaX i3D Green Premium User Manual Left blank intentionally...
Страница 161: ...15 Technical Specifications PCT 90LH User Manual 155 ENGLISH Maximum Rating Charts Emission Filament Characteristics...
Страница 166: ...15 Technical Specifications 160 PaX i3D Green Premium User Manual Left blank intentionally...
Страница 189: ...16 Appendices PCT 90LH User Manual 183 ENGLISH...
Страница 204: ......