background image

 

Installation manual

23 

 

2.  Please take out the wires, ONE72A, ONE74A and ONE75A, from the Cephalometric unit through 

the guide hole on the vertical frame unit when 

A

 and 

are close enough each other. 

But leave wire ONE73A out from this work, which will be connected to the LAN Hub later. 

 

ONE72A, ONE74A and ONE75A are 

concerned with connection this time 

 

 

 

 

 

 

 

3. Upon combining 

A

 part with 

B, 

fasten these two units using 2 wrench bolts (M8*60+8

Φ

 flat 

washer/spring]: Accessory No.15) and 1 wrench bolt(M8*50+8

Φ

 flat washer/spring]:Accessory 

No.16) at rear column. 

 

 

 

 

 

 

 

 

 

 

4.  Please fix the Column top cover (Accessory No.02) onto the back side of column, using wrench 

bolts (M3*8 and M4*8)(Accessory No. 17 and 18) 

 

 

 

 

 

 

 

 

 

 

M8*50

M8*60

M3*8

M4*8

Содержание PaX-Reve3D

Страница 1: ......

Страница 2: ......

Страница 3: ...cabinet 26 3 5 Wiring each devise and finishing installation 26 3 6 Balancing the equipment 30 Chapter 4 Installing the control box 31 Chapter 5 Technical specifications 37 5 1 Unit technical specifi...

Страница 4: ...6 5 3 4 Cephalometric Image Detector 46 5 4 Standard Accessories 47 5 5 Image Viewer programs 47 5 5 1 3D Image Viewer Ez3D2009 Standard 47 5 5 2 3D Image Viewer Ez3D2009 Professional 49 5 5 3 3D Imag...

Страница 5: ...this manual for the users to keep better comprehension of their meaning Make sure that you fully understand them and obey the instructions they contain Symbol Message meaning Note This symbol indicat...

Страница 6: ...re to ground the system 7 If there is any doubt on operation or condition do NOT install the system until a E WOO customer support team confirms the reliability Guidelines for Protection against Radia...

Страница 7: ...a wall and operator uses the system on left Width of Door The width of door is more than 800mm for system movement into X ray room Ground The condition of ground should be balanced for system balance...

Страница 8: ...for a dental diagnosis and treatment like Implant let alone the basic panoramic mode 5 Features capability of switching automatically the sensing mode even without mounting and unmounting sensors 6 E...

Страница 9: ...Installation manual 7 1 2 General view of the PaX Reve3D...

Страница 10: ...tton when a severe fault occurs Its primary use is to protect humans and equipment from a severe injuries and damages z CHIN REST help the patient rest while imaging z X RAY SENSOR Module used to dete...

Страница 11: ...and stabilize the whole equipment 1 3 Exposure switch The exposure switch enables you to launch a radiological image acquisition via exposure button from outside the X Ray room You press and hold the...

Страница 12: ...tion manual z When repairs are made by manufacturer s engineers and or trained engineers from the manufacturer z When user uses authorized components or approved components 2 2 Parts list supplied Par...

Страница 13: ...3D 1 CASE Handle table 3 Case Wire 1 1 BLOCK Column Support Block A 2 Column Support Block B 2 Base Support Block Rear fixing Plate 1 CAP Hole Cap Silicon 3 Silicon Cap 30 WRENCH BOLT 8 25 4 8 25 With...

Страница 14: ...nitor delivery status Please check whether the Shock Watch and Tilt Watch on each carton have been damaged before opening the cartons Shock Watch Tilt Watch Warning Please check the color status of Sh...

Страница 15: ...D is to be installed is flat and dry Back Left Right Front At least 3 people are needed to carry and install this unit to prevent this product from being damaged and person from being injured Please k...

Страница 16: ...14 Procedures 1 Carefully unpack the carton and position the unit as follows That is lay the unit on the floor with the column down for preparing installation 2 Please remove the moving handle using...

Страница 17: ...10 fully inside the lower part of column as the arrows indicated below UP Down Underside of lower part of column Rear bracing plate for the base 4 Please assemble the column and the base together Be c...

Страница 18: ...quired to do this Think about the unit is too heavy Firmly screw 4 wrench bolts M8 25 Accessory No 11 as stated in the following figure Important one person should grasp the bar pulling the unit towar...

Страница 19: ...rom the unit 7 By factory default to prevent the unit from being damaged while shipping and moving the unit the rotating unit and vertical frame are firmly fixed to a particular position using wrench...

Страница 20: ...PaX Reve3D 18 8 Please connect the power supply and Up Down switch cables 9 Turn on the unit and move up the column unit by approximately 20cm using the Up Down switch Power supply UP DOWN...

Страница 21: ...o work on fastening the frontal part of assembly of column and base Two procedural works are required to fix column and base together with 6 wrench bolts screws M8 25 Wrench bolt 8 Flat washer spring...

Страница 22: ...embly into the column and move it up fully so that one end of it falls in the hole Figure 2 Next pull it down through the hole and match the end of it onto the hole on the base Figure 3 Do this proced...

Страница 23: ...Installation manual 21 Figure 3 Figure 4 12 Please cover the open area where work has just been done with the base cover in a way that two magnets face down on the base Magnets...

Страница 24: ...sor is so susceptible to the physical impacts 1 Do follow these procedures to mount the Cephalometric unit on the column unit Insert the supplied Guide Shaft Assembly into the column and move it up fu...

Страница 25: ...e connected to the LAN Hub later ONE72A ONE74A and ONE75A are concerned with connection this time 3 Upon combining A part with B fasten these two units using 2 wrench bolts M8 60 8 flat washer spring...

Страница 26: ...PaX Reve3D 24 5 Please install the Case Column Top Assembly Accessory No 05 onto the front side of column Magnets...

Страница 27: ...cables from the Touch panel screen through the wire guide hole as indicated in red circle The touch pad screen is supplied with Accessory No 09 2 Then attach the touch pad screen onto the column and...

Страница 28: ...ttach the cabinet after adjusting two bolts on the side of the column 3 5 Wiring each devise and finishing installation 1 Connecting wires to the connector panel on the back of column Please connect t...

Страница 29: ...s components Connect these various cables ONE72A ONE75A ONE74A and ONE54A to each corresponding component according to the following diagram Always be careful not to plug forcefully 3 The ONE73A cable...

Страница 30: ...to those from the column For these works always be careful to not force too much when plugging each other Double check before wiring together and follow the schematic diagram Cable number Cabling name...

Страница 31: ...Top using 6 truss bolts M4 8 Accessory No 20 6 Next fasten the Case column power Accessory No 06 using 4 truss bolts M4 8 Accessory No 20 Figure 1 And fix the Column Back Case cover Accessory No 02 t...

Страница 32: ...er quality of image The level is required to do this job 1 First place the level on the front top edge of the vertical frame as shown in the figure While observing the level adjust the bolts on both r...

Страница 33: ...or the LCD panel and its control box assembly LCD panel optional control box exposure switch 2 Installing control box with the basic configuration Locate the place where control box is to be mounted a...

Страница 34: ...connector used to connect control box to the PaX Reve3D B is a 6 pin connector to provide the interoperability of this unit and other equipments making X ray exposure possible with single control box...

Страница 35: ...Installation manual 33 Combine the control box with bracket that has already been installed on the wall And screw them with 4 truss bolts M4 10 Works have just finished...

Страница 36: ...ptional Locate the place where control box is to be mounted and screw wall mounted type bracket using kal block and tapping crews M4 10 Match the corresponding holes of the control box and the rear br...

Страница 37: ...the interoperability of this unit and other equipments making X ray exposure possible with single control box SPK provides communications between the patient and operator through speaker They are phy...

Страница 38: ...PaX Reve3D 36 Combine the control box with bracket that has already been installed on the wall And screw them with 4 truss bolts M4 10 Works have just finished...

Страница 39: ...s Standard Panoramic Adult Child 13 5 sec 12 0 sec Hemi Panoramic Left and Right 6 8 sec Frontal Dentition 10 6 sec TMJ Open Close mouth 11 2 sec 4 2 8 sec Maxillary Sinus 11 0 sec Cephalometric Exami...

Страница 40: ...n Area Panoramic Mode 300 x 148 mm z Pixel Resolution 10 00 lp mm z Gray Level 14 bits 3 Cephalometric imaging detector z Technology Flat panel detector z Active Area 264x 325mm z Pixel Resolution 3 9...

Страница 41: ...Installation manual 39 5 1 4 Dimension of beam limiting device...

Страница 42: ...PaX Reve3D 40 5 1 5 Dimension of the Unit Unit mm...

Страница 43: ...41 5 1 6 Focal spot distance FOD ODD FDD mm Mode FOD ODD FDD Focal Spot to Object Distance Object to Detector Distance Focal Spot to Detector Distance CT 424 3 254 5 678 8 PANO 456 6 149 2 605 8 CEPH...

Страница 44: ...designed and produced to meet the following standards EN 60601 1 EN 60601 1 3 EN 60601 2 7 EN 60601 2 28 EN 60601 2 32 EN 60601 1 2 EN 61000 3 2 EN 61000 3 3 EN ISO 9001 EN ISO 13485 5 1 8 Marks Grap...

Страница 45: ...V z Tube current 1 22 mA z Focal spot 0 5 mm z Inherent filtration 0 8 mm Al z Added filtration 2 0 mm Al z Total filtration 2 8 mm Al z Filament characteristics 3 5 4 9V 3 5A max filament current z A...

Страница 46: ...10mA z High frequency generator constant potential micro processor controlled z Ripple 5 5 z Inverter frequency 36 kHz push pull z Nominal power less than 1 3 KW z High voltage DC z Exposure time Pan...

Страница 47: ...5sec 31 0 3 0 3 0 3 High 24 sec 22 Normal 15sec 16 D 80X60 0 2 0 2 0 2 High 24 sec 30 Normal 15sec 17 0 3 0 3 0 3 High 24 sec 18 Normal 15sec 12 I 50X50 0 2 0 2 0 2 High 24 sec 18 Normal 15sec 13 0 3...

Страница 48: ...8 200 Normal 15 sec I 50X50 0 2 0 2 0 2 High 24 sec 30 5 252 252 252 Normal 15 sec 0 3 0 3 0 3 High 24 sec 9 04 168 168 168 Normal 15 sec z Frame rate 30 fps z Gray scale 14 bit z Active area 144 mm 2...

Страница 49: ...s and three dimensional animating work by loading DICOM format CT image and more z Easy conversions through various rendering method as VR Volume Rendering MIP minIP X ray z More accurate 3 D image by...

Страница 50: ...terest ROI Region of Interest Segmentation Draw Mask 3D Picker Mask Overlay Output Capture Pane Region Full screen Print CINE player z Task MPR Rotating axis Move rotate adjust thickness Oblique slice...

Страница 51: ...ollowing value added functionality with all functionality at EZ3D2009 Professional z Auto Canal Drawing z Knowledge Info z Auto Cross Sectional z Tooth Detachment 5 5 4 Database File Server 2D Image V...

Страница 52: ...ee Draw Line Poly Line Measure Tool Distance Continuous Distance Angle Image Processing Invert Sharpen Film Effect Tool Image Analysis Magnifier Slide Profile Implant Simulation Add Implant Add Implan...

Страница 53: ...be reproduced transmitted or transcribed without the expressed written permission of the manufacturer and authors of this manual If you do not properly set the device setting causing the device to mal...

Страница 54: ......

Страница 55: ......

Страница 56: ...PaX Reve3D 52...

Отзывы: