PaX-i3D Green (PHT-60CFO) User Manual
153
Appendices
2.3
X-ray Scatter Dose
Test Equipment Information
Instrument
Manufacturer
Model
S / N
Dose Meter
Radcal
9015
91-1431/9913,4
PANO Mode (Xmaru1501CF)
Test Condition
Mode Tested
PANO HD / Adult
Applied Tube Voltage Peak [kVp]
80
Applied Tube Current [mA]
16
PANO HD 13.5 s [mR/hr]
1 m
(3.3 ft)
1.5 m
(4.9 ft)
2 m
(6.6 ft)
0
0.070
0.053
0.041
45
0.074
0.055
0.039
90
0.066
0.055
0.045
135
0.325
0.170
0.089
180
0.374
0.162
0.071
225
0.251
0.171
0.117
270
0.193
0.106
0.058
315
0.085
0.072
0.060
Mode
Direction [°]
[PI3DG_130U_44A_en]User Guide.indd 153
2016-05-24 오후 4:19:23
Содержание PaX-i3D Smart
Страница 1: ...Version 1 5 0 PHT 60CFO User Manual E n g l i s h...
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 10: ......
Страница 14: ......
Страница 42: ...This page is left intentionally blank...
Страница 110: ......
Страница 126: ......
Страница 127: ...Troubleshooting 9...
Страница 133: ...Disposing of the Unit 11...
Страница 146: ......
Страница 167: ...13 Samsung 1 ro 2 gil Hwaseong si Gyeonggi do Korea Postal Code 18449...