PaX-i3D Green (PHT-60CFO) User Manual
106
7 Acquiring CEPH Images
H. Click the
READY
button after the patient has been properly positioned. No
X-Ray will be emitted at this point.
I. Proceed to section
7.3 Initiating X-Ray Exposure.
7.2.5
CARPUS
Attaching the CARPUS plate
A. Match the notch on the CARPUS plate
to the end of nasal positioner.
B. Slide the CARPUS plate onto the nasal
positioner.
C. Slide the locking bolt on the CARPUS
plate towards the nasal positioner and turn
it to secure the CARPUS plate in place.
D. Ensure that the CARPUS plate is firmly
in place.
Positioning the Patient
A. Ask the patient to put his/her right hand
flat on the CARPUS plate. It is important to
ensure that the patient does not bend his/
her fingers.
B. Have the patient close their eyes and
remain still until scanning is complete.
Nasal positioner
Locking bolt
Fixed bolt
Notch
CARPUS Plate
[PI3DG_130U_44A_en]User Guide.indd 106
2016-05-24 오후 4:19:07