PaX-i3D Green (PHT-60CFO) User Manual
117
8 Acquiring CT Images
Positioning the Patient
A. Insert the bite block into the Chinrest
receptacle on the equipment.
B. Place a new hygienic cover over the
bite block.
WARNING
The hygiene cover for the bite block is intended for single
use only. Always replace the hygiene cover for each new
patient.
CAUTION
Disinfect the Chinrest and bite block using an alcohol-
based cleaning solution and wipe away all residue with dry
cloth before each new patient.
C. Loosen the temple supports by clicking
the Temple Support Wheel Adjustment
button.
D. Guide the patient —facing the Chinrest — to the equipment.
E. Adjust the height of the column using the column up/down button or switch
(optional) until the patient’s chin rests comfortably on the Chinrest.
Bite block
Hygiene cover
Lock /
release knob
for bite block
[PI3DG_130U_44A_en]User Guide.indd 117