PaX-i3D Green (PHT-60CFO) User Manual
149
Appendices
[CARPUS]
Figure
Hard (kVp / mA)
Normal (kVp / mA)
Soft (kVp / mA)
One shot
Scan
Fast
Scan
One shot
Scan
Fast
Scan
One shot
Scan
Fast
Scan
Man
60/6
60/6
60/10
60/5
60/5
60/9
60/4
60/4
60/8
Woman
60/6
60/6
60/10
60/5
60/5
60/9
60/4
60/4
60/8
Child
60/6
60/6
60/10
60/5
60/5
60/9
60/4
60/4
60/8
* Tube voltage and current variation range in the CEPHALOMETRIC mode: 60 ~ 100 kVp, 4 ~ 14 mA
C. CBCT
Xmaru1215CF master plus / Xmaru 1524CF master plus
Mode
High Resolution
Green
kVp
mA
kVp
mA
Man
Hard
95
10.0
95
10.0
Normal
95
9.0
95
9.0
Soft
95
8.5
95
8.5
Woman
Hard
95
9.5
95
9.5
Normal
95
8.5
95
8.5
Soft
95
8.0
95
8.0
Child
Hard
95
9.2
95
9.2
Normal
95
8.2
95
8.2
Soft
95
7.7
95
7.7
NOTE
Depending on the circumference of the patient’s head, X-ray intensity may be
classified as Hard, Normal, or Soft :
Soft ≤ Normal ≤ Hard
Age Group
Average head
circumference
Range (cm)
classification of head
circumference
Child
53
±
3 cm
> 53
±
3
Hard
53
±
3
Normal
53
±
3 <
Soft
Adult
56
±
3 cm
> 56
±
3
Hard
56
±
3
Normal
56
±
3 <
Soft
NOTE
- Maximally allowed tube voltage / current: kVp ± 10 % / mA ± 20 % according to
IEC60601-2-7.
- Due to image optimization performed prior to shipping, equipment data may differ
slightly from those specified in the table.
Gender
Sensor
Type
[PI3DG_130U_44A_en]User Guide.indd 149
2016-05-24 오후 4:19:22
Содержание PaX-i3D Smart
Страница 1: ...Version 1 5 0 PHT 60CFO User Manual E n g l i s h...
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 10: ......
Страница 14: ......
Страница 42: ...This page is left intentionally blank...
Страница 110: ......
Страница 126: ......
Страница 127: ...Troubleshooting 9...
Страница 133: ...Disposing of the Unit 11...
Страница 146: ......
Страница 167: ...13 Samsung 1 ro 2 gil Hwaseong si Gyeonggi do Korea Postal Code 18449...