2. Choosing an Installation Site
PaX-i Plus / Insight Installation Manual
11
2.
Choosing an Installation Site
2.1
Room Requirements
The location of this equipment should allow for high visibility of
the patient by the operator and the operator should be as near
to the patient as possible.
This equipment should not be installed on thick carpets for
stability reasons.
Anti-static floor materials should be used around the
equipment.
The PC monitor, emergency cut off switch and X-ray
Exposure Switch
should be installed in the vicinity of the
operator so that he or she can manage them simultaneously in
case of emergency.
Содержание PaX-i Insight
Страница 1: ......
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 10: ...Table of Contents vi PaX i Plus Insight Installation Manual ...
Страница 31: ...2 Choosing an Installation Site PaX i Plus Insight Installation Manual 21 Wall mount Type ...
Страница 104: ...8 Setting up PC s Environment Variables 94 PaX i Plus Insight Installation Manual This page intentionally left blank ...
Страница 111: ...9 Installing Software PaX i Plus Insight Installation Manual 101 2 Click Next to start installation ...
Страница 112: ...9 Installing Software 102 PaX i Plus Insight Installation Manual 3 Click Finish to exit the wizard ...
Страница 128: ...9 Installing Software 118 PaX i Plus Insight Installation Manual This page intentionally left blank ...
Страница 134: ...Appendix A 124 PaX i Plus Insight Installation Manual Schematic Diagram Components Supplied ...
Страница 138: ...Appendix A 128 PaX i Plus Insight Installation Manual 4 Remove the side cover ...
Страница 150: ......
Страница 151: ......