![Vatech Green X Скачать руководство пользователя страница 65](http://html1.mh-extra.com/html/vatech/green-x/green-x_installation-manual_3611326065.webp)
4. Installing the X-ray Unit: Base Stand Type
Green X / X18
(Model: PHT-75CHS) Installation Manual 57
En
glis
h
4.4.2 Wooden Floor
To fix the base unit on the wood floor,
Required Tool
Hammer Drill
L = 200 mm (7.9”)
Parts and
Accessory
Wood Screws
M12 x 70 – 2 pcs
(Part No. 34)
1. Put M12x70 wood screws into holes.
2. Tighten the screws with a hammer drill.
Содержание Green X
Страница 1: ...Installation manual Model PHT 75CHS Version 1 10 English...
Страница 2: ......
Страница 8: ......
Страница 40: ...3 Before the Installation 32 Green X X18 Model PHT 75CHS Installation Manual Transportation Example...
Страница 116: ...9 Installing Software 108 Green X X18 Model PHT 75CHS Installation Manual 3 Click Finish to complete the installation...
Страница 144: ......
Страница 146: ...Appendix A 2 Green X X18 Model PHT 75CHS Installation Manual Schematic Diagram Components Supplied...
Страница 162: ...Appendix A 18 Green X X18 Model PHT 75CHS Installation Manual This Page Left Blank Intentionally...
Страница 164: ...com...