4. Installing the X-ray Unit: Base Stand Type
Green X / X18
(Model: PHT-75CHS) Installation Manual 51
En
glis
h
3. If the measured distance meets the requirement, mark the installation location
for the wall bracket with a marker.
4. Place the wall bracket in the marked location. Mark a drilling point for anchor
bolts following the bracket holes.
5. Drill the holes following the marks. Each hole must be the size of 10.5mm x
30 mm (0.41” x 1.18”).
6. Hammer down Fisher strong anchors into drilled holes.
Required
Tool
Hammer
N/A
Parts and
Accessory
Fischer strong
anchor
M8 x 30
EHS tool
EAW H
8x30
7. Slide hex bolts into the mounted anchor with washers. Tighten the bolts until
they no longer move.
Required
Tool
Torque wrench
Spanner type
Parts and
Accessory
Hex Bolt
M8 x 15
Spring washer
M8
Flat washer
M8
Содержание Green X
Страница 1: ...Installation manual Model PHT 75CHS Version 1 10 English...
Страница 2: ......
Страница 8: ......
Страница 40: ...3 Before the Installation 32 Green X X18 Model PHT 75CHS Installation Manual Transportation Example...
Страница 116: ...9 Installing Software 108 Green X X18 Model PHT 75CHS Installation Manual 3 Click Finish to complete the installation...
Страница 144: ......
Страница 146: ...Appendix A 2 Green X X18 Model PHT 75CHS Installation Manual Schematic Diagram Components Supplied...
Страница 162: ...Appendix A 18 Green X X18 Model PHT 75CHS Installation Manual This Page Left Blank Intentionally...
Страница 164: ...com...