1. Introduction
6 Green X / X18 (Model: PHT-75CHS) Installation Manual
Installation Site
The PC Monitor, Emergency stop Switch, and X-ray
exposure switch should be installed near the operator
to manage them simultaneously in an emergency.
Proper shielding of the room is essential. The installers
are responsible for verifying that all applicable radiation
safety requirements are met since the requirement
varies depending on the country. Since the requirement
varies depending on the country. It is the installer’s
responsibility to verify that all applicable radiation safety
requirements are met.
Do not install the equipment near other devices.
Do not install the equipment in an area exposed to
strong electromagnetic fields.
Do not install the equipment in an area with the risk of
an explosion.
The electrical installation of this system shall comply
with all local code requirements for electro-medical
systems: IEC 60364-7-710:2002.
It is strongly recommended that a UPS be installed
simultaneously with the equipment.
The equipment, PC, and all peripheral devices must be
well-grounded.
This equipment complies with the following
standards:
IEC60601-1-1:2005
Standard Safety Requirements for
Medical Electrical Equipment
IEC 60601-1-2:2005
Electromagnetic Interference
IEC 60601-1-3:2005
Radiation Protection
IEC 60601-1:2005
Standards for Medical Electrical
Equipment
IEC 60950-1: 2nd edition:2005
Standards for Information
Technology Equipment
IEC60601-2-7 and IEC60601-2-28
: X-ray Tube Heads
IEC60364-7-710: 2002
Local Code Requirements for
Electro-Medical System Installation
IEC 60601-1-1:2005 regulation shall be met to their full
extent for the safety of the patients, operators, and the
environment—when any person assembles or modify a
medical electrical system by combing it with other
equipment.
Any equipment not provided by VATECH can be
connected when the following standards have complied
with IEC 60950-1 and IEC 60601-1
The electrical installation shall comply with local code
requirements for electro-medical systems: IEC 60364-
7-710: 2002.
Содержание Green X
Страница 1: ...Installation manual Model PHT 75CHS Version 1 10 English...
Страница 2: ......
Страница 8: ......
Страница 40: ...3 Before the Installation 32 Green X X18 Model PHT 75CHS Installation Manual Transportation Example...
Страница 116: ...9 Installing Software 108 Green X X18 Model PHT 75CHS Installation Manual 3 Click Finish to complete the installation...
Страница 144: ......
Страница 146: ...Appendix A 2 Green X X18 Model PHT 75CHS Installation Manual Schematic Diagram Components Supplied...
Страница 162: ...Appendix A 18 Green X X18 Model PHT 75CHS Installation Manual This Page Left Blank Intentionally...
Страница 164: ...com...