Page
139
of
244
4.4.
PANO Collimator Alignment
4.4.1.
PANO Collimator Alignment
1. Run the C:\VCaptureSW\Acquisition\PANO\WidePANO\VACAP.exe file.
2. Send [SPM_PANO], [SPM_PSON], [SPM_FIPM_0004], [SPM_FISS_0001],
[SPM_FISM_0001], [SPM_COLM_0007] commands to the command window to set the
PANO sensor and collimator.
3. Send [SPM_HV___0740], [SPM_HA___1200], and [SPM_SSB_] commands in the VACAP’s
Command window, and acquire the image.
4. Run EzEval.exe file, choose EzEvaluation for the tap menu on the top left and choose
collimator viewer from the tap menu on the top right.
5. Set the threshold value concerning the following standard, and check 0010.raw to check if it
satisfies the collimator alignment standard below.
6. If the Left and Right do not meet the standards, adjust the [SPM_CCST_XXXX] command.
To check the current value, enter [SPM_CCS?] to the command window.
7. Top/Bottom should be 1~10 pixels. If it does not meet the standard, check the collimator
position and move it.
※
Collimator Alignment standard value
Collimator
Min pixels
Max pixels
Threshold
70%
Left
2
2
Right
1
2
Top
1
10
Lower
1
10
If values are changed after the collimator alignment, conduct the collimator
speed calibration. (See
4.5
)
The
threshold
70%
area is marked in red.
Set Threshold to 70 and
click the Apply button.
Left
Right
Choose the 0010raw file.
Содержание A9 PHT-30CSS
Страница 1: ...vatech A9 Technical manual Model PHT 30CSS Version 1 01 English...
Страница 10: ...Page 9 of 244 Chapter 1 Hardware Part Service Guide 1 1 vatech A9 PHT 30CSS System Block Diagram...
Страница 12: ...Page 11 of 244 1 2 1 Vertical Frame Part...
Страница 17: ...Page 16 of 244 1 2 2 Handle Unit Remove the cover from the Handle Unit as shown in the image below Handle frame...
Страница 22: ...Page 21 of 244 1 2 3 5 Generator DG 07E22T2...
Страница 23: ...Page 22 of 244 1 2 4 CEPH Unit Disassemble the CEPH Unit as shown in the image below SUB075 CEPH Sensor...
Страница 25: ...Page 24 of 244 1 3 vatech A9 PHT 30CSS Connection Diagram 1 3 1 Power and Ground Connection Diagram...
Страница 26: ...Page 25 of 244...
Страница 27: ...Page 26 of 244...
Страница 30: ...Page 29 of 244 1 3 3 Sensor Part Connection Diagram...
Страница 31: ...Page 30 of 244 1 3 4 CEPH SENSOR Part Connection Diagram...
Страница 32: ...Page 31 of 244 1 3 5 Connection Diagram Total assembly...
Страница 34: ...Page 33 of 244 By doing this you can access Ethernet Grabber Main MCU BOARD MCU028 Power Board Noise Filter...
Страница 41: ...Page 40 of 244 12 Remove the cover as shown in the image below 13 Remove the cover as shown in the image below...
Страница 43: ...Page 42 of 244 16 Loosen ten bolts as shown in the image below...
Страница 112: ...Page 111 of 244 5 Click the V Vol button Click the Ready button...
Страница 120: ...Page 119 of 244 3 2 4 Test Result Parameter LSL Measurement value USL Pass Fail Contrast 1000HU 500...
Страница 180: ...Page 179 of 244 10 Select 50 in the M box and click the center of the flat washer...
Страница 194: ...Page 193 of 244 7 1 4 Test Result Parameter LSL Measurement value USL Pass Fail LAT start position 605 625 605 625 pixel...
Страница 196: ...Page 195 of 244 7 2 4 Test Result Parameter LSL Measurement value USL Pass Fail PA Start Position 1160 1180...
Страница 200: ...Page 199 of 244 4 When the download is complete the Success button appears as shown below Click the Close button to exit...
Страница 209: ...Page 208 of 244 8 5 Appendix 8 5 1 Main MCU Board Parameters...
Страница 244: ...vatech A9 www vatech com...