Series 616
INSTALLATION
893179ED.DOCX
Edition 2020-01-22
23/56
4.6
Initial operation
4.6.1
Setup procedure
To enable the valve for position control setup steps 1 to 3 must be performed.
Setup steps
Description
1
Power up
Turn on ex 24VDC power supply of valve.
Refer to chapter «Behavior during power up» for details.
2
Interface
configuration
Refer to chapter «Interface configuration» for details.
3
Valve
configuration
Basic configurations of valve must be adapted according to
application needs. Refer to chapter «Valve configuration» for details.
4.6.2
Logic Interface configuration
Interface configuration must be adapted according to application needs.
Default configuration:
OPEN input
CLOSE input
OPEN output
CLOSE output
Not inverted
Not inverted
Open
Close
•
Functionality of digital inputs CLOSE VALVE and OPEN VALVE must be selected.
These may be configured as ‘not inverted‘ or ‘inverted‘. Default is ‘not inverted‘.
Local operation:
(‘Control Performance Analyzer’ 4.0‘)
Remote operation:
Do configuration in menu ‘Valve’ > ‘Parameters’ >
‘Interface’.
It’s not possible to do ‘Interface
configuration‘ via remote operation.
4.6.3
Valve configuration
Basic valve configuration must be adapted according to application needs.
Definition of valve plate position in case of:
•
After power up, default position is min. conductance position.
Local operation:
with ‘Control Performance Analyzer’ 4.0, refer
to chapter: «Local operation»
Remote operation:
1. Open the CPA 4.0
2. Click [Local]
3. Click [Valve] > [Parameters]
4. Click [Valve] and do the settings
5. For saving the settings Click [Save]
It’s not possible to do ‘Valve
configuration‘ via remote operation.