Chapter 5. Wideline Module Tests
38
UNITY
INOVA Solids Hardware Installation
01-999044-00 C0501
Repeat steps 1 to 5 above. Although the echo maximum may change position,
symmetry of the horns should be the same. The edges of the spectrum will alter and
become sharper. This flag (i.e.,
compul
) changes the
π
/2
pulses from single to
composite and so checks the 90° phase shifting speed of the spectrometer. Refer to
9.
Repeat steps
above for sample 00-969104-01, the sodium solid standard
sample; tune the probe to the sodium frequency. Recall the standard test parameter
by entering
rt('/vnmr/tests/Na23wl')
. The sodium solid standard FID
appears similar to the example in
.
Figure 12. Wideline SSpectrum of Malonic Acid-d
6
With
compul='n'
1600 kHz
1200
1000
800
600
400
200
0
ACQUISITION
SAMPLE
PROCESSING
FLAGS
sfrq
61.395
date
Apr 1 lb
1000.00
il
n
tn
H2
solvent
none
sb
not used
in
n
at
0.082
file
/vnmr/~
gf
not used
dp
y
np
16384
tests/H2wl
awc
not used
hs
nn
sw
2e+06
DECOUPLING
lsfid
10
compul
n
fb
1e+06
dn
H1
phfid
90.0
tof
0
dof
0
wtfile
tpwr
57
dm
n
proc
ft
SPECIAL
pw
2.2
dmm
c
fn
65536
temp
not used
p1
0
dmf
4000
p180
0
homo
n
d1
10.000
TRANSIENTS
werr
tau1
20.0
bs
not used
wexp
tau2
15.0
ss
0
wbs
nt
16
wnt
ct
0
Figure 14.
23
Na FID of Sodium Standard (NaNO
3
) Starting at Top of Echo
Содержание UNITY INOVA NMR
Страница 13: ...Introduction 12 UNITYINOVA Solids Hardware Installation 01 999044 00 C0501 ...
Страница 17: ...Chapter 1 Getting Started 16 UNITYINOVA Solids Hardware Installation 01 999044 00 C0501 ...
Страница 21: ...Chapter 2 Wideband NMR Module 20 UNITYINOVA Solids Hardware Installation 01 999044 00 C0501 ...
Страница 61: ...Chapter 6 CRAMPS Multipulse Module Tests 60 UNITYINOVA Solids Hardware Installation 01 999044 00 C0501 ...
Страница 65: ...Index 64 UNITYINOVA Solids Hardware Installation 01 999044 00 C0501 ...