4.3 Installing the Solids Cabinet
01-999044-00 C0501
UNITY
INOVA Solids Hardware Installation
29
4.
Install the Solids Relay Driver board (00-992962-01) by inserting the four screws
removed
5.
Connect the ribbon cable at J7001 and connect the power cable at J7004. Reconnect
the tune relay cable at J7301.
6.
Connect one end of the second power cable (00-993140-00) to J7003 on the Solids
Relay Driver board. Connect the other end to the +5 V rail (upper, labeled TBxx) and
GND rail (lower, labeled TB2x) on the control card cage.
4.3 Installing the Solids Cabinet
The following steps describe how to install the solids cabinet.
illustrates the open
back of the solids cabinet and the relay panel. The table shows which steps to follow for
each solids module.
When Installing
Do These Steps
Wideline
1 through 5
CRAMPS/Multipulse
5 through 10
Complete solids
1 through 10
Figure 5. Solids Cabinet, Open Back View and the Relay Panel
K7200
K7201
K7202
K7205
Fan panel
Open relay panel
Sorensen VT
controller
Low band 1 kW
amplifier
Low band
power supply
High band 1kW
amplifier
High band EHT
power supply
Inside of relay panel
Содержание UNITY INOVA NMR
Страница 13: ...Introduction 12 UNITYINOVA Solids Hardware Installation 01 999044 00 C0501 ...
Страница 17: ...Chapter 1 Getting Started 16 UNITYINOVA Solids Hardware Installation 01 999044 00 C0501 ...
Страница 21: ...Chapter 2 Wideband NMR Module 20 UNITYINOVA Solids Hardware Installation 01 999044 00 C0501 ...
Страница 61: ...Chapter 6 CRAMPS Multipulse Module Tests 60 UNITYINOVA Solids Hardware Installation 01 999044 00 C0501 ...
Страница 65: ...Index 64 UNITYINOVA Solids Hardware Installation 01 999044 00 C0501 ...