Chapter 4. Solids Cabinet Preparation and Installation
30
UNITY
INOVA Solids Hardware Installation
01-999044-00 C0501
1.
Locate the cable connected between J4072 (liquids AMT channel B out) and K5004-
2 (preamplifier housing). Remove this cable.
2.
Locate the cable (00-990626-16) connected to K7200-C in the solids cabinet.
Connect the other end to J4072 (liquids AMT channel B out).
3.
Locate the cable (00-990626-17) connected to K7202-C in the solids cabinet
Connect the other end to K5004-2 in the preamplifier housing.
4.
Connect J7022 on the Solids Relay Driver board to J26 on the AMT 1 kW amplifier
(RF Blanking, 00-958297-16).
5.
Connect harness (00-990729-00) between the
UNITY
INOVA console and the solids
cabinet as follows:
6.
Connect J7023 on the Solids Relay Driver board to J7714 on the 1 kW Hi band
amplifier (RF Blanking, 00-958298-16).
7.
Locate the cable connected between J4073 (liquids AMT channel A out) and K5004-
3 (preamplifier housing). Remove this cable.
8.
Locate the cable (00-990626-16) connected to K7201-C in the solids cabinet.
Connect the other end to J4073 (liquids AMT channel A out).
9.
Locate the cable (00-990626-17) connected to K7203-C in the solids cabinet.
Connect the other end to K5004-3 in the preamplifier housing.
10. Connect the solids cabinet to the mains (208, 220, or 240 Vac 30A). Do not switch
the instrument on at this time. Wait until the hardware installation is completed.
4.4 Setting the AMT Amplifier Main Supply
The facility voltage entering the amplifier can be set to 230 or 208 Vac. Selection is made
by connecting a jumper on the M3000 module at the appropriate taps. This is illustrated in
Standard cabinet:
Solids Relay Driver board: J7009, J7011, J7030
Solids Status board: J7101, J7103, J7104, J7106, J7107, J7108, J7109, J7110
Solids cabinet:
Solid relay panel: J7206
1 kW high-band amplifier: J7703
Figure 6. AC Input at Rear of AMT M3000 Amplifier
AC
Ground
230
208
Jumper placement
Содержание UNITY INOVA NMR
Страница 13: ...Introduction 12 UNITYINOVA Solids Hardware Installation 01 999044 00 C0501 ...
Страница 17: ...Chapter 1 Getting Started 16 UNITYINOVA Solids Hardware Installation 01 999044 00 C0501 ...
Страница 21: ...Chapter 2 Wideband NMR Module 20 UNITYINOVA Solids Hardware Installation 01 999044 00 C0501 ...
Страница 61: ...Chapter 6 CRAMPS Multipulse Module Tests 60 UNITYINOVA Solids Hardware Installation 01 999044 00 C0501 ...
Страница 65: ...Index 64 UNITYINOVA Solids Hardware Installation 01 999044 00 C0501 ...