Chapter 5. Wideline Module Tests
36
UNITY
INOVA Solids Hardware Installation
01-999044-00 C0501
System Recovery Time
This test is run with no sample in the probe.
1.
Remove the sample and retune the probe.
2.
Set the spectrometer as follows:
rof2=0 sw=2e6 np=16384 nt=1 alfa=0
.
3.
Set
pw
to the value of
pw90
.
4.
Enter
go
.
Display the resulting FID. Set the cursors such that the left-hand cursor is at the very
beginning of the FID and the right-hand cursor is at the point in the FID where the
normal noise level returns. The difference between the cursors should be less than
12 µs. Refer to
Figure 10. System Recovery Time, Sample FID Display
System recovery time = 7.1
ACQUISITION
SAMPLE
PROCESSING
FLAGS
sfrq
105.800
date
Apr 1 lb
50.00
il
n
tn
Na23
solvent
none
sb
not used
in
n
at
0.004
file
/vnmr/~
gf
not used
dp
y
np
16384
tests/Na23wl
awc
not used
hs
nn
sw
2e+06
DECOUPLING
lsfid
not used
fb
1e+06
dn
H1
phfid
90.0
bs
not used
dof
0
wtfile
ss
0
dm
n
proc
ft
SPECIAL
tpwr
57
dmm
c
fn
65536
temp
not used
pw
2.2
dmf
4000
math
f
p1
0
homo
n
d1
4.000
werr
d2
0
wexp
tof
0
wbs
nt
1
wnt
ct
0
Содержание UNITY INOVA NMR
Страница 13: ...Introduction 12 UNITYINOVA Solids Hardware Installation 01 999044 00 C0501 ...
Страница 17: ...Chapter 1 Getting Started 16 UNITYINOVA Solids Hardware Installation 01 999044 00 C0501 ...
Страница 21: ...Chapter 2 Wideband NMR Module 20 UNITYINOVA Solids Hardware Installation 01 999044 00 C0501 ...
Страница 61: ...Chapter 6 CRAMPS Multipulse Module Tests 60 UNITYINOVA Solids Hardware Installation 01 999044 00 C0501 ...
Страница 65: ...Index 64 UNITYINOVA Solids Hardware Installation 01 999044 00 C0501 ...