SECTION II - INSTALLATION

87-900-879-01 (A)

5

2-6 RS-232 communicafion description

Communication format:

  8 data bit

 no 

parity

 stop 

bit

  The baud rate is programmable via a set of jumper

between 4800 and 9600 baud. The controller is
factory-set for 9600 baud operation.

Communication protocol

The communication protocol is of the masterslave type,
where:

  Host = Master

  Eurocard controller = slave

NOTE

To send 

a  command by the serial interface, the

controller must be set to "REMOTE OFF” of operation,
otherwise 

it is only possible to read the data.

The master send to the slave a word according to the
following format:

slave
address

where:

  Slave Address: controller number requested by the

master.  In RS-232 version the address is set equal
to "001";

  Parameter Number: it points out to the controller

about which parameter the master wants to
operate, and about the kind of operation
(read/write).  It consists of:

1

2

3

4

5

0=
parameter
reading
1 =
parameter
writing

always 0

000 to 299 = commands
300 to 699 = status
700 to 999 = parameter

  Data Field Lenght: it points out the next field lenght

in characters.  In this protocol the possible
combinations are:
02 = parameter reading request
06 = parameter, writing

  Data: in this controller it can be:

“=?”: in the case of parameter reading request
"000000" or "111111": in the case of YES/NO
command or parameter writing
a six numeric character string 0 to 9
pointing out the value of a configuration parameter
or an analogic measure
a six character string pointing out an error

  Checksum: it is the sum of the string ASCII values

up to the first character of the checksum with a 8 bit
cutting-off and a conversion to a decimal number.

  CR: < OxD>

All the fields are set on the right with the "O" character
used as filling character.

When a string with an address number equal to the
card one (001 in this case) is received, the controller
answers as detailed in the following:

  in the case of a parameter request with a string

having the same size described before, but the first
character of the "Parameter Number" field set to 1.

Example for a low speed status request:

Master to Controller:
0010000202= ?097< OxD >

Controller to Master:
0011000206111111016< OxD >
where " 111111" is for low speed status ON

  in the case of a parameter writing, with a string

including the parameter value that has been set out.
Example for a low speed status ON writing:

Master to Controller:
0011000206111111016< OxD >

Controller to Master:
0011000206111111016< OxD >

parameter
number

data

check-
sum

CR

data
field
lenght

Содержание Turbo-V70

Страница 1: ...87 900 879 01 A APRIL 2002 INSTRUCTION MANUAL Turbo V70 controller Eurocard Model 969 9514 vacuum technologies...

Страница 2: ...utline 3 II Installation 2 1 General 4 2 2 J2 connector 4 2 3 J3 connector 4 2 4 J4 connector 4 2 5 J5 RS 232 interface connector 4 2 6 RS 232 communication description 5 III Operation 3 1 General 7 3...

Страница 3: ......

Страница 4: ......

Страница 5: ...operation maintenance manuals and any additional information provided by Varian All warnings and cautions should be read carefully and strictly observed Address any safety operation and or maintenance...

Страница 6: ...ncy converter which is driven by a single chip microcomputer and is composed of a PCB which includes all the circuitry necessary for its operation The microcomputer generates the variable output volta...

Страница 7: ...SECTION I DESCRIPTION 87 900 879 01 A 3 1 4 Controller outline The outline dimension for the controller are shown in Fig 1 1 Figure 1 1 Controller outline...

Страница 8: ...the following Pin 3 Failure When a fault condition is detected the pump is stopped and a signal is present on this pin Pin 5 Low Speed mode activated When the Low Speed mode is activated a signal is p...

Страница 9: ...ter reading request 06 parameter writing Data in this controller it can be in the case of parameter reading request 000000 or 111111 in the case of YES NO command or parameter writing a six numeric ch...

Страница 10: ...l to 1 can be only read equal to 1 can be only written PARAMETER NUMBER TYPE DESCRIPTION X0002 D Low speed ON OFF X0003 D Pump ON OFF X0008 D Remote ON OFF 10009 D Error acknowledgment error status cl...

Страница 11: ...the same connector To switch off the pump is necessary to connect pin 12 of J4 connector to pin IS ground of the same connector 3 3 Low speed activation deactivation To activate the Low Speed status...

Страница 12: ...signals the error occurred by means of a diagnostic LED located on the board and on the RS 232 port The LED blinks in a coded mode it flashes a number of time equal to the error code see the following...

Страница 13: ...TIFICATION Product Description Varian P N Varian S N Purchase Reference TYPE OF RETURN check appropriate box Paid Exchange Paid Repair Warranty Exchange Warranty Repair Loaner Return Credit Shipping E...

Страница 14: ...rate No zero high backround Gauge tube not working Display problem Vacuum system unstable Cannot reach test mode Communication failure Degas not working Failed to start Other Error code on display Oth...

Страница 15: ...an India PVT LTD 101 108 1st Floor 1010 Competent House 7 Nangal Raya Business Centre New Delhi 110 046 India Tel 91 11 5548444 Fax 91 11 5548445 Italy Varian Vacuum Technologies Via F lli Varian 54 1...

Отзывы: