SECTION II - INSTALLATION

87-900-879-01 (A)

6

  in the case of a string with a lenght greater than 40

characters, with:

N

A

K

CR

Controller address

  in the case of a wrong checksum, with:

N

A

K

CR

Controller address

  in the case of a parameter number not present,

with:

1 0

0 6 N O - D E F

contoll.

address

param.

address

check-

sum

CR

  in the case of an out-limit value, with:

1

0

0

6

-

R A N G E

contoll.

address

param.

address

check-

sum

CR

  in the case of an attempt of writing a logic

parameter with a value not equal to " 000000" or " I
I I I I 1", or when the parameter is a read only one,
with:.

1

0

0

6

-

L O G

I

C

contoll.

address

param.

address

check-

sum

CR

  in the case of a time between two characters

greater than 1 second, with:

N

A

K

CR

Controller address

Information obtained by the serial interface are listed in
the following table where the column "TYPE" points out
a logic variable with “D” (allowed values "000000" or
"111111") and a numeric variable with "N".

Variables with the first character of "Parameter
Number" equal to X can be either written or read.
Variables with that character equal to “1” can be only
read, equal to "1" can be only written.

PARAMETER

NUMBER

TYPE

DESCRIPTION

X0002

D

Low speed ON/OFF

X0003

D

Pump ON/OFF

X0008

D

Remote ON/OFF

10009

D

Error acknowledgment:
error status clearing

00302

D

K1 status

00303

N

Error code: error code
displaying (see the
following table)

00306

D

Normal YES/NO

00307

D

Starting YES/NO

00309

N

Output frequency (Hz)

00310

N

Pump current * 100 (A)

00311

N

Temperature (°C)

00312

N

Software version
The controller sends
"QExxxx" where "xxxx” is
the CRC16 of EPROM

00313

N

Power (W)

00314

N

Pump life (hours)

00315

N

Number of cycles

00316

N

Last cycle time (minutes)

00317

N

Software version. The
controller sends “Paxxxx”
where "xxxx" is the CRC16
of EEPROM

X0700

N

Run up time (minutes)
(read only)

X0701

N

Threshold (Hz)

X0702

D

Soft-start YES/NO

X0703

D

Reduced output power
(YES)
Full output power (NO)

Error Code Table

CODE

DESCRIPTION

00E000

No error

OOEOO1

Output current > 15 A

0OE002

Not connected pump

0OE003

Output current > 1.5 A for 15 sec.

0OE004

Bearing temperature > 60 °C

0OE005

Heat exchanger temperature > 60 °C

0OE006

Normal not reached at run up time

0OE007

Input voltage < 16 V for 5 sec.

00EOOS

Operating voltage < 10 V for 5 sec.

0OE009

Vp voltage 8V lower than no~ for 5 sec.

OOE010

Soft start ramp not ended within the
expected time

Содержание Turbo-V70

Страница 1: ...87 900 879 01 A APRIL 2002 INSTRUCTION MANUAL Turbo V70 controller Eurocard Model 969 9514 vacuum technologies...

Страница 2: ...utline 3 II Installation 2 1 General 4 2 2 J2 connector 4 2 3 J3 connector 4 2 4 J4 connector 4 2 5 J5 RS 232 interface connector 4 2 6 RS 232 communication description 5 III Operation 3 1 General 7 3...

Страница 3: ......

Страница 4: ......

Страница 5: ...operation maintenance manuals and any additional information provided by Varian All warnings and cautions should be read carefully and strictly observed Address any safety operation and or maintenance...

Страница 6: ...ncy converter which is driven by a single chip microcomputer and is composed of a PCB which includes all the circuitry necessary for its operation The microcomputer generates the variable output volta...

Страница 7: ...SECTION I DESCRIPTION 87 900 879 01 A 3 1 4 Controller outline The outline dimension for the controller are shown in Fig 1 1 Figure 1 1 Controller outline...

Страница 8: ...the following Pin 3 Failure When a fault condition is detected the pump is stopped and a signal is present on this pin Pin 5 Low Speed mode activated When the Low Speed mode is activated a signal is p...

Страница 9: ...ter reading request 06 parameter writing Data in this controller it can be in the case of parameter reading request 000000 or 111111 in the case of YES NO command or parameter writing a six numeric ch...

Страница 10: ...l to 1 can be only read equal to 1 can be only written PARAMETER NUMBER TYPE DESCRIPTION X0002 D Low speed ON OFF X0003 D Pump ON OFF X0008 D Remote ON OFF 10009 D Error acknowledgment error status cl...

Страница 11: ...the same connector To switch off the pump is necessary to connect pin 12 of J4 connector to pin IS ground of the same connector 3 3 Low speed activation deactivation To activate the Low Speed status...

Страница 12: ...signals the error occurred by means of a diagnostic LED located on the board and on the RS 232 port The LED blinks in a coded mode it flashes a number of time equal to the error code see the following...

Страница 13: ...TIFICATION Product Description Varian P N Varian S N Purchase Reference TYPE OF RETURN check appropriate box Paid Exchange Paid Repair Warranty Exchange Warranty Repair Loaner Return Credit Shipping E...

Страница 14: ...rate No zero high backround Gauge tube not working Display problem Vacuum system unstable Cannot reach test mode Communication failure Degas not working Failed to start Other Error code on display Oth...

Страница 15: ...an India PVT LTD 101 108 1st Floor 1010 Competent House 7 Nangal Raya Business Centre New Delhi 110 046 India Tel 91 11 5548444 Fax 91 11 5548445 Italy Varian Vacuum Technologies Via F lli Varian 54 1...

Отзывы: