TECHNICAL INFORMATION
82 87-900-988-01 (
B
)
The display shows:
C O N F I
G U R A T I
O N
H I
G H
S P E E D
S E T T I
N G :
X X X
H z
Where:
•
XXX
=
510 ÷ 530 Hz
Confirm the flasing digit by pressing the PUMP CURRENT
key.
L O W
S P E E D
S E T T I
N G :
X X X
H z
Where:
•
XXX
=
510 ÷ 530 Hz
Confirm the flasing digit by pressing the PUMP CURRENT
key.
S O F T
S T A R T
M O D E
S E L E C T I
O N :
X X X
where:
XXX
= YES or NO.
If YES is selected, the Soft Start mode allows the pump to
rump-up the Normal speed within ten steps.
When NO is selected, the Soft Start mode is deselected
and the rump-up of the pump will be done within 360
seconds.
The controller is factory set to YES.
NOTE
The Soft Start mode may be deselected/selected only
when the pump is stopped.
•
Press CYCLE NUMBER push-button to select YES or
NO
S O F T
S T A R T
M O D E
S E L E C T I
O N :
Y E S
•
Press the PUMP CURRENT push-button and the
display shows
P R O C E S S
G A S ?
A R G O N
Select the type of purge gas, the default is air. Argon can
be seleted if needed.
Turbo-V 3K-T