TECHNICAL INFORMATION
62
87-900-854-01 (C)
•
Enter the value by pressing the PUMP CURRENT
pushbutton, and the display shows:
R S 2 3 2
H O S T / P R I N T
S E L E C T I O N :
X X X X X
where:
XXXX
== means HOST or PRINT.
•
Select HOST or PRINT by pressing the CYCLE
pushbutton.
R S 2 3 2
H O S T / P R I N T
S E L E C T I O N :
H O S T
With the RS 232 connected, a bidirectional
communication is established by selecting HOST. Data
are sent to an external computer every time the
external computer asks for the values.
The data available are:
•
Pump/controller operating condition
•
Cycle time
•
Pump life
•
Pump temperature
•
Pump current
•
Pump voltage
•
Controller output frequency
•
Cycle number
•
R1 condition
•
R2 condition
•
Pump life and cycle # zeroing
•
Configuration parameter readings
•
Configuration parameter setting
If PRINT is selected and a printer is connected on
RS 232 line, a unidirectional communication is
established and every minute the data are sent to the
printer, even if the pump is not running.
The set of data available are:
•
Pump speed KRPM
•
Pump temperature
•
Pump current A
•
Pump power W
•
R1 condition
•
R2 condition
The controller is factory-set to HOST.
•
Confirm the selection by pressing the PUMP
CURRENT pushbutton.
•
In this way you enter into an operating phase
named "Monitor Relay Programming" described in
the following paragraph.