18
03-914786-00:3
READY Status
•
READY status is the “ground state” of the ProStar 400
AutoSampler which is initiated automatically after power UP.
Every mode can be entered from READY status, except
PRIORITY mode, which can be entered only when the
ProStar 400 AutoSampler is in RUN mode.
•
In addition, RUN PARAMETER and COOLING CONTROL
mode can be entered when the ProStar 400 AutoSampler is
in the RUN mode.
Function codes are used to enter any specific mode, except for
the RUN PARAMETER mode which can be entered directly via
the PROG/END key, and RUN mode (entered by the START
command).
Every mode uses the same function code (F,0) to exit, except for
RUN PARAMETER mode and RUN mode which are left via
PROG/END and STOP respectively.
Key Functions
STEP key; used to step through the parameter list. The STEP key is also used to
rotate the sample tray (in READY status only.).
START/STOP
Start/Stop automatic processing of samples.
HOLD/CONT
Hold/Continue countdown of analysis time. Analysis time is extended with the time
that hold is active.
PROG/END
Enables/Ends programming of parameters in the selected mode. Note that the
PROG/END key selects the RUN PARAMETER mode and enables the
programming of its parameters at the same time.
CL
Clears the value of the parameter on display.
F
Function key; used in combination with a numerical key to select or exit the
various modes.
0-9
Numerical keys; used to enter values for programmable parameters. Values are
stored when going to the next parameter (STEP) or when ending programming
(PROG/END).
Содержание ProStar 400
Страница 2: ......
Страница 28: ...2 03 914786 00 3...