O
PERATION
ProStar 400 AutoSampler
25
Priority Sample
A sample run can be interrupted to process an emergency
sample prior to the next sample of the programmed samples
series: press F,2 to enter the priority mode, press PROG/END
key and enter the number of the tray position where the priority
vial has been placed. Press PROG/END to end programming.
Return to RUN (F,0). Priority vial will be processed immediately
after the current vial. A blinking “
P
” (first position of lower display
line) indicates that a priority vial is to be/being processed. After
processing the priority vial, RUN will continue where it was
interrupted.
NOTE:
Priority mode can only be entered during sample run. It cannot be
entered during pre flush or loopfill.
Remote Control
When used with Varian ProStar or Star LC modules, the ProStar
400 is interfaced using Cable Kit Part Number 03-909126-90.
Remote control mode is not necessary to use this cable.
However, using remote control mode, the ProStar 400 will wait
for a “ready” signal from the other modules of the Star system
before starting its injection cycle. See the installation instructions
enclosed with the cable for details of these modes of operation.
When the ProStar 400 is in REMOTE CONTROL mode it can
only be controlled using the “next injection” and “next vial” inputs.
RUN is executed as programmed, but analysis time is now
infinite. Only a “next injection” input signal will initiate the next
injection from the same vial (or from the next vial when the
programmed number of injections per vial have been executed).
A “next vial” input signal will initiate the first injection from the
next vial.
With the
ProStar 400 in READY status, REMOTE CONTROL
mode is entered with F,3; display will show:
0 REMOTE CONTROL
READY