TECHNICAL INFORMATION 

NOTE 

Using the RS 485 interface, the message structure 
remains identical to the one used for the RS 232 
interface, the only difference being that the value 
assigned to the ADDRESS 

<ADDR>

 

Examples: 

 

Command: START

 

 

Source: PC 

 

Destination: Pump 

02 80 30 30 30 31 31 03 42 33 

STX ADDR 

WINDOW 

WR  ON  ETX 

CRC 

Source: Pump   
Destination: PC 

02 80 06 03 38 35 

STX ADDR ACK  ETX 

CRC 

 

Command: STOP

 

 

Source: PC 

 

Destination: Pump 

02 80 30 30 30 31 30 03 42 32 

STX ADDR 

WINDOW 

WR  OFF  ETX 

CRC 

Source: Pump   
Destination: PC 

02 80 06 03 38 35 

STX ADDR ACK  ETX 

CRC 

 

Command: READ PUMP STATUS 

 

Source: PC 

 

Destination: Pump (with address = 3) 

02 83 32 30 35 30 03 38 37 

STX ADDR 

WINDOW 

RD  ETX 

CRC 

Source: Pump (with address = 3 in stop status) 
Destination: PC 

02 

83  32 30 35 30 30 30 30 30 30  03  38 37

STX ADDR  WINDOW 

DATA (STATUS) 

ETX CRC 

 

Windows meaning 

 

WIN TYPE 

DESCRIPTION 

000 L START/STOP 

(1=START; 0=STOP) 

001 L LOW 

SPEED 

[0=OFF / 1=ON] 

008 L REMOTE/SERIAL 

CONFIGURATION 

(1=REMOTE; 0=SERIAL) 

102 

SET POINT VALUE 
(Hz) 

105 

SET POINT HYSTERESIS 
(%) [0-100] 

108 N BAUD 

RATE 

(0-4) [600-1200-2400-4200-9600] 

117 

LOW SPEED ADJUST 
(Hz) 

120 

ROTATIONAL FREQUENCY SETTING 
[Hz] (High speed adjust) 

200 N BUS 

CURRENT 

[mA] 

201 N 3PHASE 

VOLTAGE 

[Vrms] 

202 N POWER 

[W] 

203 N DRIVING 

FREQUENCY 

[Hz]: (current driving frequency) 

205 N STATUS 

[0=stop ; 2=ramp ; 3=autotuning ; 5=normal ; 
6=fail] 

206 N ERROR 

CODE 

Bit 7: motor block; 
Bit 6: short circuit; 
Bit 5: undervoltage; 
Bit 4: motor over temperature; 
Bit 3: power fail; 
Bit 2: controller overt; 
Bit 1: pump overt; 
Bit 0: spare 

211 

CONTROLLER TEMPERATURE [°C] 

216 

ENVIROMENT TEMPERATURE [°C] 

300 N CYCLE 

TIME 

[min] 

301 N CYCLE 

NUMBER 

302 N PUMP 

LIFE 

[h] 

319 A CONTROLLER 

MODEL 

323 

CONTROLLER SERIAL NUMBER 

325 

ELECTRICAL MODIFICATION LEVEL 

500 

Monitor Mode (write only) 

503 

RS485 SERIAL ADDRESS SETTING 
[0-31] 

504 

SERIAL TYPE SELECT 
(0=RS323; 1=RS485) 

 

 

23

 

87-900-978-01(D) 

Содержание HS452

Страница 1: ...HS452 HS652 Rotary Vane Pumps Model 949 9360 Model 949 9365 87 900 978 01 D JUNE 2006 I MANUALE DI ISTRUZIONI D BEDIENUNGSHANDBUCH F NOTICE DE MODE D EMPLOI GB INSTRUCTION MANUAL...

Страница 2: ...HS452 HS652 ROTARY VANE PUMPS...

Страница 3: ......

Страница 4: ......

Страница 5: ...D EXHAUST FLANGES 16 STARTING AND RUNNING THE PUMP 16 STOPPING THE PUMP 17 SAFETY RULES 17 WARNING NOTES 17 CAUTION NOTES 18 MAINTENANCE ACTIONS 18 LUBRICANTS 19 PUMP ELECTRONIC CONTROLLER 21 Technica...

Страница 6: ...dere la pompa e a non sottoporla ad urti o vibrazioni Non disperdere l imballo nell ambiente Il materiale comple tamente riciclabile e risponde alla direttiva CEE 85 399 per la tutela dell ambiente NO...

Страница 7: ...pe rare i 60 C PERICOLO Escludere sempre l alimentazione della pompa prima di com piere operazioni di manutenzione Apporre specifici cartelli di avvertenza APPARECCHIATURA IN MANUTENZIONE NON INSERIRE...

Страница 8: ...t en oder Vibrationen ausgesetzt wird Das Verpackungsmaterial ist ordnungsgem zu entsorgen Es ist vollst ndig recyclebar und entspricht der EG Richtlinie 85 399 f r den Umweltschutz ANMERKUNG Die Pump...

Страница 9: ...Anschlu an den Betrieb gewartet werden soll ist abzuwarten bis sie abgek hlt ist da ihre Oberfl che eine Temperatur von 60 C berschreiten kann GEFAHR Vor Wartungsarbeiten ist die Pumpe stets energiefr...

Страница 10: ...n choc ni aucune vibration Ne pas jeter l emballage dans la nature Le mat riel est enti rement recyclable et il est conforme la directive CEE 85 399 en mati re de protection de l environnement NOTE La...

Страница 11: ...de de fonctionnement il est indispensable de la laisser refroidir car sa temp rature ext rieure peut tre sup rieure 60 C DANGER Avant toute op ration de maintenance il est imp ratif de tou jours coupe...

Страница 12: ...en impact or shock vibration to it Do not dispose of the packing materials in an unauthorized manner The material is 100 recyclable and complies with EEC Directive 85 399 NOTE Normal exposure to the e...

Страница 13: ...uter surface may be in excess of 60 C WARNING Always disconnect the power supply to the pump before start ing maintenance work Place a special warning signs over the power supply breaker switch MACHIN...

Страница 14: ...uces pump noise The pump is equipped with a special anti suckback device which automatically isolates the vacuum system when the pump stops This avoids rises in pressure or oil flow in the vacuum syst...

Страница 15: ...uum system are completely isolated from each other and air can enter the pump without any risk for the vacuum produced in the system GAS BALLAST VALVE When the pump sucks in vapors these condense duri...

Страница 16: ...TECHNICAL INFORMATION View of the pump 11 87 900 978 01 D...

Страница 17: ...stage rotor SR03700367 44 Tie rod SR03700335 45 OR gasket 1 2 46 Bracket 47 Screw 49 Valve cover plate 50 Valve spring 1 2 51 Valve gasket 1 2 N Description HS452 652 52 First stage stator SR03700419...

Страница 18: ...h 550 550 OIL CAPACITY min max l 0 6 1 0 6 1 MOTOR RATING 3ph kW 0 5 0 5 NOMINAL ROTATION SPEED low speed 66 rpm 1980 1980 46 rpm 1350 1350 NOISE LEVEL with gas ballast closed dB A 60 60 OIL TEMPERATU...

Страница 19: ...TECHNICAL INFORMATION DIMENSIONS The following figure shows the pump dimensions HS 452 and HS 652 pump dimensions 14 87 900 978 01 D...

Страница 20: ...ould ideally be a concrete floor and of any accessories mounted on it Note that the pump is stable on its base plate and it should not be necessary to anchor it to the floor with bolts and screws also...

Страница 21: ...asing with the risk of breaking the pumping block envelope STARTING AND RUNNING THE PUMP Before being put into service in order to reach maximum vacuum the pump must be left running for about an hour...

Страница 22: ...ker switch MACHINE UNDERGOING MAINTENANCE DO NOT POWER ON c If you are performing maintenance after the pump has been operating for a considerable time allow sufficient time for it to cool as the exte...

Страница 23: ...nd unscheduled maintenance work Maintenance includes 1 SCHEDULED MAINTENANCE Maintaining the nominal state of operation 2 UNSCHEDULED MAINTENANCE Restoring the nominal state of operation NOTE The freq...

Страница 24: ...nto water or in the sewage system All violations of these rules are liable to persecution as provided for by law When using lubricants always keep an oil can nearby Take care in draining off waste oil...

Страница 25: ...e 25 C 1 10 5 Boiling point 0 01 torr C 112 Viscosity 40 C cSt SUS 55 258 Viscosity 100 C cSt SUS 8 1 52 7 Viscosity index 119 Pour point C F 15 5 Flash point C F 213 415 Fire point C F 244 472 Densit...

Страница 26: ...y contact out 2 Speed setting 0 10V in 3 Oil level relay contact out 4 Start in 5 Ground out 6 Norm Status N O relay contact out 7 Oil level relay contact out 8 Start in 9 24V out Remote Start Stop Ro...

Страница 27: ...type as per the following table DATA TYPE FIELD LENGTH VALID CHARACTERS Logic L 1 0 OFF 1 ON Numeric N 6 0 9 right justified with 0 Alphanumeric A 10 from blank to _ ASCII ETX End of transmission 0x03...

Страница 28: ...ws meaning WIN TYPE DESCRIPTION 000 L START STOP 1 START 0 STOP 001 L LOW SPEED 0 OFF 1 ON 008 L REMOTE SERIAL CONFIGURATION 1 REMOTE 0 SERIAL 102 N SET POINT VALUE Hz 105 N SET POINT HYSTERESIS 0 100...

Страница 29: ...of the oil temperature and of the chamber volume to rough down The lower is the oil temperature the longer is the time required to reach nominal speed Oil level check An oil level check is performed f...

Страница 30: ...obtain the IP44 protection level you must set up the pump electrical connections as highlighted in the following figure NOTE You must exchange the 9 pin D mating connector shells with the IP44 version...

Страница 31: ...come pericoloso EFFETTI ACUTI L esposizione alle nebbie d olio pu causare nausea e irritazione agli occhi Non sono stati eseguiti studi dettagliati tuttavia non si ritiene che il materiale possa caus...

Страница 32: ...mezzo assorbente 7 Movimentazione e magazzinaggio MOVIMENTAZIONE Nessun rischio conosciuto MAGAZZINAGGIO Nessun rischio conosciuto 8 Controlli dell esposizione MISURE DI CONTROLLO Protezione personale...

Страница 33: ...essuna INALAZIONE ACUTA Il limite degli USA di esposizione alle nebbie d olio TWA 8 ore di 5 mg al metro cubo 12 Informazioni ecologiche AMBIENTALI Non sono previsti effetti am bientali negativi nel c...

Страница 34: ...gef hrlich klassifiziert AKUTE WIRKUNGEN Sich ligen D mpfen auszusetzen kann belkeit und Augenreizungen verursachen Es wurden keine detaillierten Untersuchungen durchgef hrt es wird jedoch nicht erwar...

Страница 35: ...werden 7 Handhabung und Lagerung HANDHABUNG Nicht bekannt LAGERUNG Nicht bekannt 8 Expositionskontrolle Personenschutz TECHNISCHE KONTROLL Nicht erforderlich MASSNAHMEN ATEMSCHUTZ Siehe Anmerkungen zu...

Страница 36: ...Keine AKUTES EINATMEN Der US amerikanische Grenzwert f r die lnebelexposition TWA 8 Stunden betr gt 5 mg pro Kubikmeter 12 Umweltinformationen UMWELT Bei vorgeschriebener Verwendung Entsorgung sind ke...

Страница 37: ...comme dangereux EFFETS AIGUS L exposition aux vapeurs d huiles peut causer des naus es et l irritation des yeux Des tudes d taill es n ont pas t effectu es mais on ne s attend pas ce que le mat riel p...

Страница 38: ...Manipulation et stockage MANIPULATION Aucun risque connu STOCKAGE Aucun risque connu 8 Contr les de l exposition MESURES DE CONTR LE Aucune mesure n cessaire Protection personnelle TECHNIQUE PROTECTIO...

Страница 39: ...pour TWA de 8 heures d exposition aux vapeurs d huile est de 5 mg par m tre cube 12 Informations cologiques ENVIRONNEMENT Si le produit est utilis et ou mis au rebut de la fa on indiqu e aucun effet d...

Страница 40: ...ROUTES Ingestion inhalation of oil mists This product is not classified as hazardous ACUTE EFFECTS Exposure to oils mists may cause nausea and eye irritation Detailed studies have not been made but ma...

Страница 41: ...ith an absorbent 7 Handling and Storage HANDLING None Known STORAGE None Known 8 Exposure Controls Personal Protection ENGINEERING CONTROL MEASURES None required RESPIRATORY PROTECTION See notes on ve...

Страница 42: ...MAL LD 50 None ACUTE INHALATION US Gov t 8 hr TWA limit for exposure to oil mists is 5 mg per cubic meter 12 Ecological Information ENVIRONMENTAL When used and or disposed of as indicated no adverse e...

Страница 43: ...TIFICATION Product Description Varian P N Varian S N Purchase Reference TYPE OF RETURN check appropriate box Paid Exchange Paid Repair Warranty Exchange Warranty Repair Loaner Return Credit Shipping E...

Страница 44: ...rate No zero high backround Gauge tube not working Display problem Vacuum system unstable Cannot reach test mode Communication failure Degas not working Failed to start Other Error code on display Oth...

Страница 45: ...1263 Toll Free 0120 655 040 Korea Varian Technologies Korea Ltd Shinsa 2nd Bldg 2F 966 5 Daechi dong Kangnam gu Seoul Korea 135 280 Tel 82 2 3452 2452 Fax 82 2 3452 2451 Toll Free 080 222 2452 Mexico...

Отзывы: