DOC: 03-925880-00
DESC:
PAGE:
REV
1 of 7
3800GC, Burn-In Test Procedure
2
PREPARED BY: Sax Makthepharack REVISED BY:
CONTROL: INDUSTRIAL ENGINEERING
REV. DATE:
SHOWN
-
07 - 30 - 99
F.Faili/M.Slaughter
1.
3800GC BURN-IN PROCEDURE
1.1
TOOLS AND EQUIPMENT REQUIRED:
1.1.1
Flow meter
1.1.2
Injector nuts/seals T-8941
1.1.3
Torque wrench, 15 in/lbs with 3/32 hex tip
1.1.4
Test column, 6 ft. long (cut to three Feet long) P/N 03-912403-35)
1.1.5
#2 Philips screw driver
1.1.6
Straight Slot screw driver
1.1.7
Open end wrench, 5/16-1/4
1.1.8
Open end wrench, 7/32
1.1.9
Open end wrench, ¾
1.1.10 Plug, detector base
1.1.11 Packed Column adapter, 03-918631-00
1.1.12 Adapter Seal, 03-925132-00
1.1.13 Voltmeter
1.1.14 Capillary Pick
1.1.15 Ground Strap
1.1.16 Gases, N2, He, Air, H2
1.1.17 Injector Nut wrench
1.1.18 Tweezers
1.1.19 Hy-Pot Tester
1.1.20 Quad fitting, 1/8” (with 3 pieces of peek tubing & manifold plate)
1.2
Verify G.C. to S/O & Specials Documentation if Applicable
1.3
Hy-Pot Instrument (observe safety precautions)
1.3.1
Display Safety Chain in front of Instrument.
1.3.2
Attach Ground Cable to the Instrument Frame.
1.3.3
Display a Red Safety Cone on top of the unit being tested..