DOC: ATP-03-925110-XX
DESC:
PAGE:
REV
10 of 16
TEST & CALIBRATION
2
PREPARED BY: HOWIE PLITSCH REVISED BY:
CONTROL: INDUSTRIAL ENGINEERING
REV. DATE:
SHOWN
-
08 - 01 - 00
Sax Makthepharack
General Description of Functions
No Function
- Default
Serial # and Config Changer
- Only use the Serial Number function of this utility. Other settings to be set by the Manufacturing Engineer.
*
Set Function to "Change Det Serial Number" or "Change Inj Serial Number"
*
Input the new Serial Number in the "New EFC Seriall Number" block.
*
Press "Run"
*
Change Function to "No Function"
*
Press Exit
Exit
- Press Run and you will exit the EFC Calibration program.
Test and Calibrate
- Routine used after Quicktest to create and write calibration data to memory of EFC modules.
*
Fixture # is indicated on Fixture.
*
Reference Flow Sensor Number is indicated on Flow Sensor
*
Type and Revision reflect the configuration and PWA revision respectively.