990 CLD Autoline™ Leak Detector
DR
AFT 2/2/04
1-11
7
HIGH VACUUM PUMP
O
Pluggable terminal block for connections
to the Connector high vacuum pump.
See “Wire Run Lists” on page 2-4.
8
OPTO-22 Modules
N/A
Input voltage selection modules for
filament and remote exponent control
signals.
9
General I/O Connector
I/O
Pluggable terminal block for connections
to the equipment as described below.
9-1 (+), 9-2 (
−
)
FILAMENT ON
I
Voltage applied for at least two seconds
causes the active filament in the
spectrometer tube to light if the unit is in
the Remote Filament Control mode.
9-3 (+), 9-4 (
−
)
FILAMENT OFF
I
Voltage applied for at least two seconds
causes the active filament in the
spectrometer tube to shut off if the unit is
in the Remote Filament Control mode.
See.
9-5 (+), 9-6 (-)
PRESSURE
O
Spectrometer tube pressure monitor
output, 0 to + 10 V. 4 V = filament
protection threshold.
9-7 (+), 9-8 (-)
LEAK RATE
O
Leak rate monitor output, 2 or 3
V/decade logarithmic, or 0 to 10 V
linear.
9-9, 9-10, 9-11
PRESSURE THRESHOLD
O
Relay energizes when spectrometer tube
pressure is too high for operation.
9-13,9-14
FILAMENT STATUS
O
Relay contact closes when the active
filament in the spectrometer tube is on.
9-16, 9-17
HIGH VACUUM PUMP
O
Relay contact closes when high vacuum
pump is STATUS functional and
operating at its correct speed.
Table 1-6
Rear Panel Inputs and Outputs (Continued)
Item Description
Input/
Output
Function
Note
The input voltages applied to the General I/O connector depend on the optional
OPTO-22 input module purchased. These modules are available in 2.5 to 28
VDC., 35 to 60 VDC, and 120 VAC (5“Parts List” on page 5-1).