990 CLD Autoline™ Leak Detector
DR
AFT 2/2/04
1-9
NOTE
Pin connections read from left to right as viewed from the rear
of the console.
Table 1-5
Secondary Operating Controls and Indicators
Item Control/Indicator
Function
1
SENSITIVITY HI/LOW
Toggle switch that changes the speed of the high vacuum pump.
It is used to change the sensitivity of the leak detector by a factor
of ten.
2
CALIBRATE
Potentiometer used to adjust the gain of the leak detector's
amplifiers. Used to calibrate the leak detector.
3
EMIS
Screwdriver-adjustable potentiometer adjusts the level of
electron emission current in the spectrometer tube ion source.
4
AUDIO VOL/THRESH Concentric potentiometers. The outer knob varies the tone's
volume heard at the speaker. The inner knob varies the audible
threshold of a given leak rate.
5 FILAMENT
OFF
-
ON/AUTO
Toggle switch to either manually turn off the filament or place it
in the ON/AUTO mode.
6 FILAMENT,1,
FILAMENT 2
Toggle switch which selects the active filament in the
spectrometer tube ion source.
7
COARSE ZERO
Potentiometer used to zero any background signal.
8
CHECK/RUN
When the toggle switch is in the CHECK position, the leak
detector is rendered insensitive to helium. When in the RUN
position, the leak detector is sensitive to helium.
9
ION
Ten-turn ion voltage adjustment potentiometer used for tuning
the leak detector to helium.
10
FOCUS
Potentiometer used to adjust the focus voltage of the ion beam.